enabled.
How to configure digital filter:
You can configure the digital Filter either through the front panel
operation or through the remote interface operation.
Front Panel Operation
For MODE setting:
1. Press SHIFT + DIGITS buttons and then use
between
O FF
"
2. Press ENTER on your selection.
For READINGS setting:
1. After enabling the filter (see above), use
"READINGS". Press ENTER to select it. Use
through the digits and
numbers to a desired value(from 2 to 100). Press ENTER to set the
value. Press ENTER on "RUN" to start reading.
Remote Interface Operation
Use the following commands to set your digital filter:
SENSe:AVERage:TCONtrol {MOVing|REPeat}
SENSe:AVERage:TCONtrol?
SENSe:AVERage:COUNt {<value>|MINimum|MAXimum}
SENSe:AVERage:COUNt? [MINimum|MAXimum]
SENSe:AVERage:STATe {OFF|ON}
SENSe:AVERage:STATe?
4.1.3 Resolution & NPLC Setting
Definition
Resolution is the number of digits a multimeter can measure. You can
select the resolution for a specific measurement. The choices for the
resolution setting are: 4.5, 5.5 and 6.5. For a faster measurement, you
can select 0.001 PLC. For a stable, higher accuracy measurement, 100
, MOVING AVG
"
"
and
a nd
and
R EPEAT AVG
"
a nd
buttons to increase or decrease the
48 48 48 48
b uttons to toggle
.
"
b uttons to locate
and
to move