Ac Filter (Ac Only) - Picotest M3510A User Manual

6 1/2
Table of Contents

Advertisement

4.1.2.1 AC Filter (AC only)

Definition:
You can refer to Table 4-2 to set the bandwidth for selecting one of the
three AC filters (Slow, Medium and Fast), in order to achieve either
higher accuracy in low frequency measurements or faster AC settling
time.
Defaults
The factory default is 20 Hz (Medium). You can select your filter type as
you wish when setting bandwidth value. The defined selection is stored in
a volatile memory and the default setting will be restored when the
meter is power-off.
Bandwidth
3 Hz ~ 300 KHz
20 Hz ~ 300 KHz
200 Hz ~ 300 KHz
How to set AC Filter in AC measurements:
User can set the AC Filter either through the front panel operation or
through the remote interface operation.
Front Panel Operation
1. Press ACV button and then CONFIG button.
2. Use
and
press ENTER to select it. (There are three options: 3Hz, 20Hz and
200Hz.)
Remote Interface Operation
From your PC terminal, use the following command to specify the filter
type:
Table 4-2
AC Filter
Slow
Medium
Fast
buttons to locate "BAND WIDTH" submenu, and then
46 46 46 46
Remote Trigger Delay
( seconds per reading)
7
1
0.6

Hide quick links: