Download Print this page
GE LOGIQ 9 Basic Service Manual
Hide thumbs Also See for LOGIQ 9:

Advertisement

Quick Links

P R E L I M I N A R Y
GE Healthcare
LOGIQ™ 9
Basic Service Manual
1/18/10
GEHC_FRNT_CVR.FM
Part Number: 2294853-100
Revision: 13

Advertisement

loading
Need help?

Need help?

Do you have a question about the LOGIQ 9 and is the answer not in the manual?

Questions and answers

Summary of Contents for GE LOGIQ 9

  • Page 1 P R E L I M I N A R Y 1/18/10 GEHC_FRNT_CVR.FM GE Healthcare LOGIQ™ 9 Basic Service Manual Part Number: 2294853-100 Revision: 13...
  • Page 3 • WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN WARNUNG (DE) ANWWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
  • Page 4 P R E L I M I N A R Y 1/18/10 PRECAUTIONS_AND_REV_A GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL...
  • Page 5 P R E L I M I N A R Y 1/18/10 PRECAUTIONS_AND_REV_A GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL VAROITUS (FI) • ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ...
  • Page 6 P R E L I M I N A R Y 1/18/10 PRECAUTIONS_AND_REV_A GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL • DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK. • HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
  • Page 7 P R E L I M I N A R Y 1/18/10 PRECAUTIONS_AND_REV_A GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL • DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK. • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å...
  • Page 8 P R E L I M I N A R Y 1/18/10 PRECAUTIONS_AND_REV_A GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL • DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA. • OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA...
  • Page 9 P R E L I M I N A R Y 1/18/10 PRECAUTIONS_AND_REV_A GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL (JA) (ZH-CN) (KO)
  • Page 10 GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 11 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 12 P R E L I M I N A R Y 1/18/10 PRECAUTIONS_AND_REV_A GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Revision History Date Revision (YYYY/MM/DD) Reason for change 2002/04/08 Initial Release 2002/11/08 Update for BT’02 Release, BEP and FEP changes 2003/04/04 Update for R2.1.1 Release and hardware changes...
  • Page 13 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
  • Page 14 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Dangerous Procedure Warnings ........1 - 18 Battery Safety .
  • Page 15 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 2 Site Preparation Overview............2 - 1 Purpose of Chapter 2 .
  • Page 16 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 3 System Setup Overview ............3 - 1 The Purpose of Chapter 3 .
  • Page 17 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Available Probes........... 3 - 21 Software/Option Configuration .
  • Page 18 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose of Chapter 4 .
  • Page 19 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL For BT03 and prior ........4 - 18 For BT04 and later .
  • Page 20 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview ............5 - 1 Purpose of Chapter 5 .
  • Page 21 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Location of the Back End Processor ......5 - 24 Inputs (BEP 1) .
  • Page 22 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Back End Processor (BEP 4.x) ........5 - 39 Enclosure .
  • Page 23 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Top Console ............5 - 63 Monitor .
  • Page 24 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Logs ..........5 - 83 Utilities .
  • Page 25 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 6 Service Adjustments Overview............6 - 1 Purpose of Chapter 6 .
  • Page 26 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Tools ............6 - 13 Manpower .
  • Page 27 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
  • Page 28 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL PC Diagnostics ..........7 - 31 Non-Interactive Tests - BT04 and earlier .
  • Page 29 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview............8 - 1 Purpose of Chapter 8 .
  • Page 30 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Save Connectivity Settings ........8 - 17 DICOM TCPIP Settings .
  • Page 31 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL To Place the System in the Standby Mode ....8 - 68 To Wake the System from Standby Mode .
  • Page 32 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 9 Overview ............9 - 1 Purpose of Chapter 9 .
  • Page 33 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Peripheral Cables ..........9 - 44 Peripheral Cable - Analog Color Printer with BEP3 .
  • Page 34 Safety Test Overview ......... . 10 - 11 GE Healthcare Leakage Current Limits for LOGIQ™ 9 ....10 - 12 Outlet Test - Wiring Arrangement - USA &...
  • Page 35 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chassis Leakage Current Test ........10 - 15 Definition .
  • Page 36 P R E L I M I N A R Y 1/18/10 L9_TOC.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL xxxiv Table of Contents...
  • Page 37 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ™ 9 scanner. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
  • Page 38 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 1-2 Service Manual Overview 1-2-1 Contents in this Service Manual Chapter 1 - Introduction: Contains a content summary and warnings.
  • Page 39 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-2-3 LOGIQ™ 9 Models Covered by this Manual Table 1-2 LOGIQ™ 9 Model Designations...
  • Page 40 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-2-4 Product Description 1-2-4-1 Overview of the Product LOGIQ™ 9 is a phased and linear array ultrasound imaging scanner. It has provisions for analog input sources like ECG and Phono.
  • Page 41 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-2-4-1 Overview of the Product (cont’d) Name Plate Monitor (CRT) Monitor Controls Op. Panel Upper Bezel...
  • Page 42 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-2-4-1 Overview of the Product (cont’d) 17”LCD Monitor Monitor Controls Name Plate Op. Panel Upper Bezel...
  • Page 43 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-2-4-1 Overview of the Product (cont’d) 19” LCD Monitor Monitor Controls Name Plate Op. Panel Upper Bezel...
  • Page 44 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-2-4-8 Product Manuals NOTE: The LOGIQ™ 9 Unpacking Procedure is delivered on paper. Section 9-23 "Customer Documentation"...
  • Page 45 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 1-3 Important Conventions 1-3-1 Conventions Used in Book 1-3-1-1 Model Designations This manual covers the LOGIQ™ 9 scanners.
  • Page 46 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-3-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter.
  • Page 47 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-3-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment.
  • Page 48 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION Defibrillator-proof type of equipment. At the ECG connector on front of system.
  • Page 49 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION "Mains ON" Indicates the power on position of the mains power switch.
  • Page 50 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION This symbol indicates that waste electrical and electronic equipment must not be disposed of as unsorted municipal waste and must be collected separately.
  • Page 51 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION How to lower LCD prior to transport Rear of the LCD monitor.
  • Page 52 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 1-4 Safety Considerations 1-4-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment.
  • Page 53 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-4-3 Mechanical Safety (cont’d) The LOGIQ™ 9 weighs at least 171kg (376 lbs.) when ready to use, depending on installed CAUTION monitors and peripherals.
  • Page 54 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 1-4-4-2 Probes Follow these guidelines before connecting a probe to the scanner: • Inspect the probe prior to each use for damage or degradation to the:...
  • Page 55 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 1-6 Dangerous Procedure Warnings (cont’d) When the top console is in its locked position, the gas shock is compressed and...
  • Page 56 If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during use, recharging or storage, immediately remove it and stop using it. If you have any questions about the battery, consult GE or your local representative. •...
  • Page 57 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 1-8 Lockout/Tagout Requirements (For USA Only) Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the electrical Mains plug.
  • Page 58 P R E L I M I N A R Y 1/18/10 CHAPTER1.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 1-10 Electromagnetic Compatibility (EMC) 1-10-1 What is EMC? Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment.
  • Page 59 Service 1-800-321-7937 Latin America Applications Support 1-262-524-5698 Europe Phone: +33 (0) 130-831-300 (General Imaging and Cardiac) GE Ultraschall Deutschland GmbH& Co. KG +43 (0) 7682-3800-26 (Voluson-Logiqbook) BeethovenstraBe 239 Fax: +49 (0)2122-8024-31 Postfach 11 05 60, D-42655 Solingen Germany Asia (Singapore)
  • Page 60 IRECTION EVISION ASIC ERVICE ANUAL 1-11-2 System Manufacturer Table 1-7 System Manufacturer Manufacturer FAX Number GE Healthcare - GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC 9900 Innovation Drive 414-721-3865 Wauwatosa, WI 53226 1-24 Section 1-11 - Customer Assistance...
  • Page 61 P R E L I M I N A R Y 1/18/10 CHAPTER2.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 2 Site Preparation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ 9.
  • Page 62 2-2-2 Electrical Requirements NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 63 2-2-2-5 Unit Power Plug If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
  • Page 64 Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit RF RF gaskets touch metal leakage. Or, if a label has been found in such a position, move the label. Use GE specified harnesses The interconnect cables are grounded and require ferrite beads and other shielding. Also, cable and peripherals length, material, and routing are all important;...
  • Page 65 P R E L I M I N A R Y 1/18/10 CHAPTER2.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 2-2-4 Scan Probe Environmental Requirements Table 2-5 Environmental Requirements - Probes Operating Temperature Storage Temperature Standard Probes -10°...
  • Page 66 (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 67 2-3-2 Required Features NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 68 P R E L I M I N A R Y 1/18/10 CHAPTER2.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 2-3-4 Minimal Floor Plan Suggestion Figure 2-1 Minimal Floor Plan Section 2-3 - Facility Needs...
  • Page 69 P R E L I M I N A R Y 1/18/10 CHAPTER2.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 2-3-5 Networking Setup Requirements 2-3-5-1 Stand Alone Scanner (without Network Connection) None. 2-3-5-2 Scanner Connected to Hospital’s Network...
  • Page 70 P R E L I M I N A R Y 1/18/10 CHAPTER2.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 2-10 Section 2-3 - Facility Needs...
  • Page 71 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 3 System Setup Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to unpack and set up the unit. Included are procedures to receive, unpack and configure the equipment.
  • Page 72 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-2 Set Up Reminders 3-2-1 Average Set Up Time Table 3-2 Average Set Up Time...
  • Page 73 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH...
  • Page 74 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment Two people are needed to unpack the unit because of its weight.
  • Page 75 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 3-4 Unpacking Procedure (Continued) Step Task Press once on the swivel lock to release the brake.
  • Page 76 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-4 Preparing for Set Up 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered, or damaged.
  • Page 77 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-4-2-3 ETL Testing Laboratories Safety Rating Label This ETL label is required for all USA systems.
  • Page 78 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-5 Completing the Setup 3-5-1 Power On / Boot Up NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 79 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-5-1-2 Back-end Processor Boot Up 1.) Press the On / Off switch on the Control Panel once.
  • Page 80 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-5-2-2 Scanner Shutdown Cable Clip Circuit Breaker Mains Power Cable Figure 3-7 Circuit Breaker and Power Cable on Back of Scanner, with Power On 1.) Switch OFF the Circuit Breaker at the back of the system.
  • Page 81 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions The physical dimensions of the LOGIQ™ 9 unit are summarized in...
  • Page 82 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-6-3 On-Board Optional Peripherals Table 3-8 Approved On-Board Peripherals Device Manufacturer Model Video Signal...
  • Page 83 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-6-4 External I/O Connector Panel (Back of system) Figure 3-8 External I/O Panel Labels (Prior to BT03)
  • Page 84 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-6-4 External I/O Connector Panel (Back of system) (cont’d) Table 3-9 External I/O Connector Descriptions...
  • Page 85 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-6-4-1 External I/O Pin Outs Table 3-10 Footswitch Connector Pin No Output Signal Description...
  • Page 86 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-6-4-1 External I/O Pin Outs (cont’d) Table 3-14 S-VGA Connector, Shrinked Sub-D 15 Pin...
  • Page 87 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-6-5 Other I/O Ports (Front of system) Figure 3-10 Front Connectors Table 3-17 Front I/O Connector Descriptions...
  • Page 88 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-6-5 Other I/O Ports (Front of system) (cont’d) Table 3-18 CW Connector, J1 (Optional)
  • Page 89 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-6-5 Other I/O Ports (Front of system) (cont’d) Table 3-21 Flock of Birds, Large Connector (Optional, not shown in Figure 3-10)
  • Page 90 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 3-6-6 Video Specification Video specifications may be needed to be able to connect laser cameras or other devices to the LOGIQ™...
  • Page 91 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-7 Available Probes See Specifications in the LOGIQ™ 9 User Reference Manual for probes and intended use.
  • Page 92 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-9 Connectivity Setup Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ SN:...
  • Page 93 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-10 Before Starting a Software Upgrade This information has been moved, see Section 8-6 "Before Starting a Software Load" on page 8-17...
  • Page 94 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-14 Connecting USB Line Printers - Black/White and Color 3-14-1 Printer Set Up (R4.x and later) 1.) Attach the USB cable from the printer to the USB port on the back of the ultrasound system.
  • Page 95 User Manual(s) Check that the correct User Manual(s), for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 System Setup...
  • Page 96 P R E L I M I N A R Y 1/18/10 CHAPTER3.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 3-16 Checks Perform all functional checks. 3-26 Section 3-16 - Checks...
  • Page 97 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ 9 scanner diagnostics by using the built-in service software, and power supply adjustments.
  • Page 98 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 4-2 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
  • Page 99 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-1 System Features 4-2-1-1 Control Panel Figure 4-1 Control Panel Tour 1. Touch Panel and Controls 11.
  • Page 100 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-1-2 Touch Panel Figure 4-2 Touch Panel Display Tour - BT06 Table 4-3 Touch Panel Display 1.
  • Page 101 15. Imaging Parameters by Mode. 2. Patient Name, Patient Identification. 16. Focal Zone Indicator. 3. Power Output Readout 17. TGC. 4. GE Symbol: Probe Orientation Marker. 18. Depth Scale. 5. Image Preview. 19. Body Pattern. 6. Gray/Color Bar. 20. Image Management Icons: Active Images, Delete, Previous/Next 7.
  • Page 102 OPERATE THIS UNIT ONLY WHEN ALL BOARD COVERS AND FRAME PANELS ARE SECURELY IN PLACE. THE COVERS ARE REQUIRED FOR SAFE OPERATION, GOOD SYSTEM PERFORMANCE AND COOLING PURPOSES. Use only power supply cords, cables, and plugs provided by or designated by GE Healthcare. NOTICE NOTE: Do not cycle the Circuit Breaker ON-OFF-ON in less than five -5- seconds.
  • Page 103 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-2 Power On / Boot Up (cont’d) 7.) Connect the Power Cable to an appropriate mains power outlet, allowing sufficient slack so that the unit can be moved slightly.
  • Page 104 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-3 Standby Mode Use Standby Mode to decrease the power off/on time while transporting system for a portable exam.
  • Page 105 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-4 Power Off / Shutdown NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 106 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-4-2 To Shut Down Scanner Cable Clip Circuit Breaker Mains Power Cable Figure 4-7 Circuit Breaker and Power Cable on Back of Scanner...
  • Page 107 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-5 Probe/Connectors Usage 4-2-5-1 To Connect a Probe 1.) Place the probe's carrying case on a stable surface and open the case.
  • Page 108 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-6 Using CINE 4-2-6-1 To Activate CINE 1.) Press Freeze, then roll the Trackball to activate CINE.
  • Page 109 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-7 Image Management Quick Guide / User Guide For Image Management functionality, refer to the LOGIQ™ 9 Quick Guide / User Guide. It talks about several topics: •...
  • Page 110 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-8 System B/M-Mode Checks For a basic functional check of the system’s different modes, the Quick Guide / User Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
  • Page 111 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-9 System CFM and PWD Checks Figure 4-9 CFM and PWD Functions Table 4-7...
  • Page 112 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-10 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze.
  • Page 113 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-10-3 Worksheets Measurement/Calculation worksheets are available to display and edit measurements and calculations. There are generic worksheets as well as Application specific worksheets. The worksheets are selected from the Measurement Touch Panel.
  • Page 114 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-11 The MOD (Magnetic Optical Drive) 4-2-11-1 For BT03 and prior The onboard 3.5 inch Magnetic Optical disk drive supports the following densities: Standard: 640MB;...
  • Page 115 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-11-2 For BT04 and later The LOGIQ™ 9 supports the portable, external Fujitsu DynaMO 1300 U2 Pocket Magnetic Optical (MO) drive to read old media and to use up unused MO disks.
  • Page 116 If you intend to use the MO drive semi-permanently, attach the MO drive to the system using velcro, or other secure method (not supplied by GE). DO NOT lock the Control Panel into place while the MO Drive is located above the Front Peripheral Storage Area.
  • Page 117 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-2-11-3 Using the MOD 1.) Before installing an MO disk in the MOD, check the MO disk for loose hardware or damaged labels which could jam inside the MOD.
  • Page 118 NOTE: Do not create multiple partitions. Use USB Drive with single partition. The USB Flash Drive is an ESD-sensitive device. GE only supports the NDL 256MByte 066E0690 USB Flash Drive. Before removing the USB Flash Drive from the USB port, press Eject (F3) and select USB Drive from CAUTION the pull-down menu to disconnect the USB Flash Drive.
  • Page 119 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 4-3 Software Configuration Checks Table 4-8 Software Configuration Checks Step Task to do...
  • Page 120 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 4-4 Peripheral Checks Check that peripherals work as described below: Table 4-9 Peripheral Checks...
  • Page 121 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 4-5 Mechanical Function Checks 4-5-1 Operator I/O Movement - Standard Monitor Figure 4-13...
  • Page 122 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-5-2 Operator I/O Movement - 17” LCD (Option) Hold here for Steps 1 and 2...
  • Page 123 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-5-3 Operator I/O Movement - 19” LCD (Option) Hold here for Steps 1 and 2...
  • Page 124 P R E L I M I N A R Y 1/18/10 CHAPTER4.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 4-5-4 Brakes and Direction Locks Checks Check that the brakes and direction locks function as described below.
  • Page 125 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQ™...
  • Page 126 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-2 General Information LOGIQ™ 9 is a phased and linear array ultrasound imaging scanner. It has provisions for analog input sources like ECG and Phono.
  • Page 127 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-2 General Information (cont’d) 17”LCD Monitor Monitor Controls Name Plate Op. Panel Upper Bezel...
  • Page 128 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-2 General Information (cont’d) 19” LCD Monitor Monitor Controls Name Plate Op. Panel Upper Bezel...
  • Page 129 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-2-1 Block Diagram BACK END PROCESSOR USER INTERFACE ECG Patient Interface AC POWER 3.5" MOD 17"...
  • Page 130 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-2-2 Dataflow Control Description 5-2-2-1 Data Sets Each scan mode will produce data to a corresponding data set in the Image Memory. The data sets are represented in ring-buffers.
  • Page 131 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-3 Front End Processor The Front End Processor generates the strong bursts transmitted by the probes as ultrasound into the body.
  • Page 132 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-1 Front End Processor Power Supply Board (FEPS, FEPS2, FEPS2.1) The power supply assembly receives approximately 220VAC from the secondary side of the mains isolation transformer.
  • Page 133 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-3 Front End Subsystem The LOGIQ™ 9 Front End Subsystem consists the XDIF (Transducer Interface) board, RF Amp board and eight TD (Transmit Delay) boards.
  • Page 134 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-3-3 RF Amp, RF Amp1.2, RF Amp2 Board (Radio Frequency Amplifier Board) Figure 5-8 RF Amp, RF Amp2 Functions...
  • Page 135 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-3-4 TD, TD4, TD5.x, TD6 (Time Delay) Boards (8) Figure 5-10 Basic TD Board Functions...
  • Page 136 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-3-4 TD, TD4, TD5.x, TD6 (Time Delay) Boards (8) (cont’d) Figure 5-12 Basic TD Input/Output Control Signals Major control signal from the EQ board are the Digital TGC/VREF and Serial Control Bus (I C).
  • Page 137 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-4 Scan Control Board (SCB, SCB2) The Scan Control Board combines onto one board the basic functionality of Image Port (IP), Front End Control and Timing (FECT), Scan Sequencer (SS2) and the Scan Trigger (System Timing).
  • Page 138 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-4-1 SCB, SCB2 High Level Features • Generates and buffers system clocks: 40 MHz (40P0 and 40P1)
  • Page 139 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-4-3 Scan Control Board LEDs (cont’d) Table 5-4 SCB2 2365739 LED Indications Number LED Color...
  • Page 140 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-5 Equalization (EQ) Board (component of EBM Board) 5-3-5-1 General Description The EQ dynamically gains and filters the received signal output from the final beamformer sum. TGC control calculates range dependent analog and digital gains.
  • Page 141 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-5-2 Basic Functions • Compensates for the attenuation of the transmitted signal in tissue by depth: performs TGC based on focal zone.
  • Page 142 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-6 B/M Mode Processor (BMP) Board 5-3-6-1 General Description The B/M-mode Processor (BMP) subsystem takes the digital complex data from the EQ subsystem and converts it to either B-mode (2D) or M-mode (timeline) data.
  • Page 143 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-7 EBM, EBM2 Board The EBM Board was used on forward production units starting with BT’02 (November 2002). The EBM2 Board was used on forward production units starting with R3.0.0 Software (BT’03 October 2003).
  • Page 144 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-7 EBM, EBM2 Board (cont’d) Power Supply Con Power Supply Con HV Line HV Line...
  • Page 145 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-3-7 EBM, EBM2 Board (cont’d) 5-3-7-4 B–Mode Output Blocks The B–mode output block collect vector data and tags for delivery to the Scan Control Board. It can form output vectors from one or multiple input vectors (multi–transmit focus).
  • Page 146 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-4 Back End Processor (BEP 1) The Back End Processor (BEP 1) receives the data from the Front End Electronics, stores it in memory, performs scan conversion to pixel domain, and drives the system RGB monitor.
  • Page 147 Full Direct 3D and Open-GI support (desirable). 5-4-4 PCIO Board This GE designed board does not plug into any slot but is mounted to the inside of the PC chassis. Power and Internal PCI control signals 5-4-5 PCI Video Card...
  • Page 148 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-4-6 UPS Battery Operation The UPS batteries provide power to the Back End Processor to enable a controlled shut down of the processor.
  • Page 149 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-4-9 Inputs (BEP 1) 5-4-9-1 AC Power Table 5-5 AC Power Input Description Connection from:...
  • Page 150 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-4-10 Outputs Table 5-9 Output Signals Signal Name Description Signal Path +5VDC DC Voltage PC2IO >...
  • Page 151 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-5 Back End Processor (BEP 2) The Back End Processor (BEP 2) receives the data from the Front End Electronics, stores it in memory, performs scan conversion to pixel domain, and drives the system RGB monitor.
  • Page 152 • Full Direct 3D and Open-GI support (desirable). 5-5-4 PCIO Board: This GE designed board does not plug into any slot but is mounted to the inside of the PC chassis. Power and Internal PCI control signals 5-5-5 PCVIC Card Input from Motherboard AGP Video (BEP 2.0)
  • Page 153 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-5-6 DGVIC, DGIO and ADD Cards BEP 2.1 (BT’02, November 2002) The ADD Card is a PCI card that mimics the AGP video and provides a digital video output to the DGVIC board.
  • Page 154 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-5-11 Location of the Back End Processor BEP 2 Rear Side Front of of Scanner...
  • Page 155 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-5-12 Inputs 5-5-12-1 AC Power Table 5-10 AC Power Input Description Connection from: 230 VAC AC Power AC Distribution Box >...
  • Page 156 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-5-12-4 Bi-directional Signals Table 5-13 Bi-directional Signals Signal Name Description Signal Path UPS Control Control Signals to and from the UPS BEP >...
  • Page 157 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-5-13 Outputs Table 5-15 Output Signals Signal Name Description Signal Path +5VDC DC Voltage DG2IO >...
  • Page 158 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-6 Back End Processor (BEP 3.x) The Back End Processor (BEP 3.x) receives the data from the Front End Electronics, stores it in memory, performs scan conversion to pixel domain, and drives the system RGB monitor.
  • Page 159 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-6-1 Enclosure The EMC enclosure shall house an ATX Power Supply with UPS battery backup, an ATX style motherboard with minimum of 512 MB of RAM and a 3.2 GHz processor.
  • Page 160 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-6-6 Internal Storage Devices: • A 80 Gigabyte EIDE Hard Disk Drive (HDD) inside the Back End Processor cabinet.
  • Page 161 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-6-8 Inputs (BEP 3.x) 5-6-8-1 AC Power Table 5-16 AC Power Input Description Connection from:...
  • Page 162 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-6-9 Outputs Table 5-20 Output Signals Signal Name Description Signal Path +5VDC DC Voltage Bulkhead >...
  • Page 163 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-7 Back End Processor (BEP 4.x) The Back End Processor (BEP 4.x) receives the data from the Front End Electronics, stores it in memory, performs scan conversion to pixel domain, and drives the system RGB monitor.
  • Page 164 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-7-1 Enclosure The EMC enclosure shall house an ATX Power Supply with UPS battery backup, an ATX style motherboard with minimum of 1 GB of RAM and a 3.4 GHz processor.
  • Page 165 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-7-6 Internal Storage Devices: • A 80 Gigabyte EIDE Hard Disk Drive (HDD) inside the Back End Processor cabinet.
  • Page 166 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-7-8 Inputs (BEP 4.x) 5-7-8-1 AC Power Table 5-21 AC Power Input Description Connection from:...
  • Page 167 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-7-9 Outputs Table 5-25 Output Signals Signal Name Description Signal Path +5VDC DC Voltage Bulkhead >...
  • Page 168 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-8 BEP Power On Path (refer to Figure 5-26 on page 5-44 The Back End Processor is powered ON by: Press and release the OP I/O Standby (On/Off) switch.
  • Page 169 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-9 BEP Power Off Path (refer to Figure 5-27 on page 5-45 The Back End Processor is powered Off by: Press and release the OP I/O Standby (On/Off) switch.
  • Page 170 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-10 Operator Console Control PWA Figure 5-29 Figure 5-28 show the audio path through the Operator Console Control Printed Wire Assembly.
  • Page 171 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-10-1 Console Main Cable 5-10-1-1 Main Cable Connector Functions - Prior to BT04 Figure 5-30 Main Cable Drawing - prior to BT04...
  • Page 172 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-10-1-1 Main Cable Connector Functions - Prior to BT04 (cont’d) Table 5-26 Main Cable Composition Prior to BT04...
  • Page 173 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-10-1-1 Main Cable Connector Functions - Prior to BT04 (cont’d) Table 5-29 Connector C & H - Touch Screen Interface...
  • Page 174 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-10-1-1 Main Cable Connector Functions - Prior to BT04 (cont’d) Table 5-33 Connector K- Op Console Power, Audio, Mic, Control...
  • Page 175 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-10-1-3 Main Cable Connector Functions - BT04 and after Figure 5-31 Main Cable - BT04 and after...
  • Page 176 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-10-1-4 Main Cable Connectors with Pinouts - BT04 and after (5119218) Table 5-36 Connector A & G- Safety Ground - BT04 and after (5119218)
  • Page 177 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-10-1-4 Main Cable Connectors with Pinouts - BT04 and after (5119218) (cont’d) Table 5-39 Connector D &...
  • Page 178 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-11 Patient I/O (Currently not supported by the LOGIQ™ 9) The Patient I/O is mounted inside the Backend Processor’s cabinet with its connector panel available on the front of the scanner.
  • Page 179 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-12 Internal I/O The Internal I/O is the interface between the Card Cage, PC Module, and the rest of the system, including the Top Console and the Peripherals.
  • Page 180 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-12 Internal I/O (cont’d) TO CARD RACK SONY VCR SVO-9500MD Remote Right AUDIO OUT...
  • Page 181 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-12 Internal I/O (cont’d) Figure 5-35 Internal I/O Back End Processor Connections The Internal I/O board performs buffering and distribution of video and control data from various sources to multiple destinations.
  • Page 182 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-12-1 Location in the Unit The Internal I/O Module is plugged into the Card Rack (at the end nearest to the rear of the system).
  • Page 183 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-12-5 Inputs 5-12-5-1 DC Voltages Table 5-42 DC Input Voltages Input Description Connection from +3.3 VDC...
  • Page 184 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-12-5 Inputs (cont’d) 5-12-5-4 Serial Ports Table 5-45 Serial Ports Input Description Connection from...
  • Page 185 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-12-6 Outputs 5-12-6-1 DC Voltage Table 5-47 DC Output Voltages Output Description Connection to...
  • Page 186 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-12-6 Outputs (cont’d) 5-12-6-4 Interface Connectors – Component Side Table 5-50 Connectors, Component Side...
  • Page 187 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-13 Top Console The Top Console includes a Stand By/On switch, a keyboard, different controls for manipulating the picture quality, and controls for use in Measure &...
  • Page 188 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-14 Monitor 17-inch High Resolution Monitor with no interlace scan includes loudspeakers for stereo sound output (used during Doppler scanning).
  • Page 189 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-15 External I/O The External I/O is the interface between the scanner and all external items.
  • Page 190 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-15-3 Bi-directional Signals Table 5-52 Bi-directional Signals Signal Name Description Signal Path Ethernet - External I/O - Internal I/O -...
  • Page 191 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-15-4 Output Signals Table 5-53 Output Signals Signal Name Description Signal Path Audio Source (Internal Stereo audio to external VCR.
  • Page 192 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-16 Peripherals A DVR or VCR, an analog or digital Black-and-White Video Printer, and an analog or digital Color Printer may be installed on the scanner.
  • Page 193 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-17 Modem and USB Modem 5-17-1 General Description Multi-Tech global modem. The modem is a standard modem that connects to an analog phone line for InSite capabilities.
  • Page 194 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-17-6 LEDs (non-USB Modem) The Modem has 10 LEDs on its front. TDRD CD 56 3314OHTRECFX Figure 5-43 LEDs on Modem’s Front Panel...
  • Page 195 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-18 Power Distribution Figure 5-44 Overall Power Distribution Chapter 5 Components and Functions (Theory)
  • Page 196 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-18-1 AC Power Distribution Overview The AC Power’s main tasks are to supply the various internal subsystems with AC power and to galvanically isolate the scanner from the on site Mains Power System.
  • Page 197 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-18-2 AC Power PWB (cont’d) Input voltage to the transformer can be either 230VAC or 115 VAC (two different part numbers/FRUs).
  • Page 198 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-18-3 AC Controller Differences Figure 5-47 Original AC Controller Functions 5-74 Section 5-18 - Power Distribution...
  • Page 199 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-18-3 AC Controller Differences (cont’d) Figure 5-48 AC Controller FC 200403/FC200404 Chapter 5 Components and Functions (Theory)
  • Page 200 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-19 Mechanical Descriptions 5-19-1 Monitor 5-19-1-1 Standard • Tilt: 10° forwards and backwards.
  • Page 201 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-19-2-1 17” LCD (Option) • Tilt Adjustable Monitor: Tilt Angle: Up 10°, Down 90°...
  • Page 202 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-19-3 Wheels/Brakes Check that the brakes and direction locks function as described below. Table 5-55...
  • Page 203 The Service Platform contains a set of software modules that are common to all ultrasound and cardiology systems containing a PC backend. This web-enabled technology provides linkage to e-Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever. The Service Platform will increase service productivity and reduce training and service costs.
  • Page 204 Global Service User Interface (GSUI) 5-21-2-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. 5-21-2-2 Service Login Select the wrench icon in the status bar at the bottom of the scan display screen.
  • Page 205 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-21-3 Service Home Page The navigation bar at the top of the Customer Service Home Page screen allows the user to select from several tools and utilities.
  • Page 206 However, the basic functionality has not changed. Also, some additional navigational features have been added, such as Last Page and Get Page. The Filter Error log is keyed to GE Service login access permissions and is not available to customer level analysis.
  • Page 207 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-21-4-1 Logs The eight sub-menus of the Logs category are System, Power, Infomatics, Temperature, Probe, Board, DICOM, and UPS Power.
  • Page 208 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-21-4-2 Utilities The two sub-menus of the Utilities category are Plot Log, and Plot Page.
  • Page 209 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-21-4-4 Exit The sub-menu, Exit Log Viewer, returns the user to the Service Desktop home page.
  • Page 210 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-21-5 Diagnostics Detailed Diagnostic information is found in Chapter 7 - Diagnostics/Troubleshooting. NOTE: After performing any diagnostics, detach probes and reboot the system.
  • Page 211 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-21-6 Image Quality The Image Quality page is intended to contain tools for troubleshooting image quality issues.
  • Page 212 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-21-8 Configuration The Configuration page is intended to be used to setup various configuration files on the system.
  • Page 213 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-21-9 Utilities The Utilities page contains several miscellaneous tools such as System Shutdown, the Image Viewer Utility, and the iLinq Configuration Utility.
  • Page 214 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-21-11 Planned Maintenance (PM) The PM page is intended to contain the tools used in periodic maintenance of the system.
  • Page 215 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-22 4D Option Beginning with BT04 (Application Software R4.X.X), the LOGIQ™ 9 can be configured to perform 4D scans.
  • Page 216 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-22-2 4D Block Diagram Figure 5-65 4D BLock Diagram 5-22-2-1 Motor Controller to Front Bracket Cable and 4D Probe Cable The connection between the motor control box and the front bulkhead is a 15 pin sub-D (at the motor controller) and a LEMO (at the bulkhead).
  • Page 217 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-22-2 4D Block Diagram (cont’d) Pin Diagram of 7 pin Lemo connector Pin Diagram of 12 pin Lemo connector on the small end of the 4D Probe Cable on the large end of the 4D Probe Cable.
  • Page 218 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-22-2 4D Block Diagram (cont’d) The pinout of the motor controller to the front bracket requires a 15 pin SUB-D female Connector (MC4)
  • Page 219 ERVICE ANUAL 5-22-2 4D Block Diagram (cont’d) The LOGIQ 9 3D/4D motor control box also includes a power supply with connection to the L9 110 VAC power distribution (MC1). 5-22-2-3 Motor Controller to Internal I/O Cable A 9 pin SUB-D connection to the L9 Internal IO(MC3) requires the pin out as shown below.
  • Page 220 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-23 Continuous Wave Doppler Option Beginning with BT04 (Application Software R4.X.X), the LOGIQ™ 9 can be configured to perform Continuous Wave (CW) Doppler scans.
  • Page 221 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 5-24 PCI CW Doppler PWA card The CW Doppler has a local controller that configures transmit and receive paths. The controller performs all required signal processing in order to produce 16 bit signed integer I/Q data path.
  • Page 222 P R E L I M I N A R Y 1/18/10 CHAPTER5.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 5-24-3 Mu Metal Shield Mu metal shield is used to block flux field from the L9 isolation transformer mounted below the BEP.
  • Page 223 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification.
  • Page 224 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 6-3 Monitor NOTE: After performing any diagnostics, detach probes and reboot the system.
  • Page 225 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-3-1-2 Brightness and Contrast Table 6-2 Brightness and Contrast - Standard Room Condition Contrast...
  • Page 226 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-3-1-3 Visible Image Display Adjustment If you get a monitor where the image is smaller than the physical CRT displayed area, in other words the image does not fill the screen, you can adjust the size of the visible image display: 1.) Press and HOLD the center button "brightness/contrast (icons)"...
  • Page 227 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-3-2-5 Procedure 1.) Raise the Control Panel to a height that allows you to work underneath the monitor.
  • Page 228 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-3-3 Monitor Adjustment (17” LCD Option Only) 6-3-3-1 Brightness and Contrast Table 6-3 Brightness and Contrast - 17” LCD...
  • Page 229 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-3-4 Monitor Adjustment (19” LCD Only) 6-3-4-1 Brightness and Contrast Table 6-5 Brightness and Contrast - 19” LCD...
  • Page 230 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 6-4 Panel Calibration The Touch Panel Adjustment/Calibration procedure can be found in Section 7-6-1-1 “Calibration Utilities”...
  • Page 231 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-5-4 Z-Release Cable Adjustment Before making any adjustments, check the Z-release system to determine what adjustments CAUTION need to be made.
  • Page 232 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-5-4-2 Cable Adjustment 1.) From underneath the handle use a 10mm open-end (small “ignition” style) wrench to screw in the cable adjustment screw to tighten the Z-release cable.
  • Page 233 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-5-5-2 Procedure NOTE: The operator panel must be at or near its highest position.
  • Page 234 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-5-6 Emergency Release Procedure When the operator panel is at or near its highest position it is possible to see the gas spring mechanism with the Black and White Printer or the VCR removed.
  • Page 235 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 6-6 Brake Arm Adjustment Adjustments to the brake arm may be necessary if the pedals do not depress readily nor engage so the front wheels lock in place and do not swivel.
  • Page 236 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-6-4 Brake Arm Adjustment Procedure To adjust the brake arm it is recommended to separate both front adjustment nuts from the brake assembly.
  • Page 237 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 6-7 Brake Adjustment 6-7-1 Front Wheel Brakes Service adjustment to the front brakes involve setting the brake so it will properly “bite” the rear tire. The default setting should provide proper braking.
  • Page 238 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-7-2 Rear Wheel Brakes Adjustment Service adjustment to the rear brakes involves setting the brake so it will properly “bite” the rear tire.
  • Page 239 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 6-7-2 Rear Wheel Brakes Adjustment (cont’d) 2 screws removed Area on either side 2 screws removed...
  • Page 240 P R E L I M I N A R Y 1/18/10 CHAPTER6.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 6-18 Section 6-7 - Brake Adjustment...
  • Page 241 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 242 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 7-2 Service Safety Considerations DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN DANGER THIS EQUIPMENT.
  • Page 243 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 7-3 Gathering Trouble Data 7-3-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition through remote diagnostics (InSite) or to be sent back to the manufacturer for analysis.
  • Page 244 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-3-3 Collecting a Trouble Image with Logs If the system malfunctions, press the Alt-D keys simultaneously. This Alt-D function is available at all times, and collects a screen capture of the image monitor, user-defined presets, and the following logs: •...
  • Page 245 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-3-3 Collecting a Trouble Image with Logs (cont’d) NOTE: Restart the application before resuming clinical scanning.
  • Page 246 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-3-4 Capturing Network Logs with Network Sniffer A sniffer monitors network traffic and allows you to capture network data without redirecting or altering it.
  • Page 247 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-3-4 Capturing Network Logs with Network Sniffer (cont’d) NOTE: Steps describe how to filter the data so that you only record this system’s network activity.
  • Page 248 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-3-4 Capturing Network Logs with Network Sniffer (cont’d) 4.) On the Select Filter Task screen (Figure 7-6), select “Show only IP packets FROM/TO one or more...
  • Page 249 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-3-4 Capturing Network Logs with Network Sniffer (cont’d) 7.) In the Filter to Apply field, select “My Computer”.
  • Page 250 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-3-4 Capturing Network Logs with Network Sniffer (cont’d) 12.)When the transaction finishes (or fails), select Ctrl+Alt+Del to open Task Manager.
  • Page 251 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-3-4 Capturing Network Logs with Network Sniffer (cont’d) 14.)On the Distinct Network Monitor screen (Figure 7-3), select Capture >...
  • Page 252 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-3-4 Capturing Network Logs with Network Sniffer (cont’d) 17.)On the Save As Capture File Dialog (Figure 7-11), select SAVE.
  • Page 253 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 7-4 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center through InSite.
  • Page 254 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-4-1 Check and Record the P4 Key Function (cont’d) P4 is the factory default Screen Capture Key. If it is not set to Whole Screen or Screen Capture, as...
  • Page 255 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-4-3 Capturing a Screen The following is a generic process to capture any screen from the scanner: 1.) Navigate to and display the image/screen to be captured.
  • Page 256 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-4-3 Capturing a Screen (cont’d) Figure 7-15 Save Dialog Box 6.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the image on the hard disk, MOD or CD-R.
  • Page 257 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-4-5 Ctrl-PrintScreen Shortcut Starting with R4.x.x software, a Ctrl-PrintScreen shortcut is available for quickly capturing the image displayed on the system.
  • Page 258 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 7-5 Diagnostics NOTE: Detach probes and reboot system after performing any diagnostics. As described in...
  • Page 259 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 7-6 Acquisition Diagnostics NOTE: After performing any diagnostics, detach probes and reboot the system.
  • Page 260 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 7-6 Acquisition Diagnostics (cont’d) 4.) Loop Count This is an editable text field that will only accept numeric values of 4 digits or less. When the switch is configured as an “execute”...
  • Page 261 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-6-1 Diagnostic Listings Table 7-2 Acquisition Diagnostics and Utility Listing Diagnostic/Utility Function SCB PCI Interface Diag Tests the host’s ability to communicate with the SCB via the secondary PCI bus...
  • Page 262 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 7-2 Acquisition Diagnostics and Utility Listing (Continued) Diagnostic/Utility Function Set TGC Utility Set the analog TGC to a specified value.
  • Page 263 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-6-1-1 Calibration Utilities Touchscreen Calibration: Follow the directions on the Touchscreen. As each of the cross-hairs appear, touch them with your finger or a pencil eraser.
  • Page 264 2.) Locate the Op Panel Test under I/O Devices > Op Panel Utilities > Op Panel Interface. 3.) Select Start to launch the GE Test App. An hour glass displays on the Service Platform window. The GE Test App window is hidden behind the Service Platform window.
  • Page 265 Diagnostic Listings (cont’d) Attach Button Active Detach and Other Buttons Active Figure 7-19 Attach and GE Test App Startup Tab 6.) Select the tabs to test the various features of the Op Panel. 7-7-1 Trackball Tab Test the trackball functionality by moving the pointer over the blue box. See Figure 7-20.
  • Page 266 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-7-2 Keyboard Tab Test the keyboard functionality by typing on the keyboard. Verify that the corresponding key is highlighted when each key is pressed.
  • Page 267 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-7-4 Encoders Tab Test the encoder functionality by rotating each encoder knob. See Figure 7-23.
  • Page 268 ERVICE ANUAL 7-7-6 LEDs Tab Test the LED functionality by selecting each button or knob on the GE Test App. See Figure 7-25. Verify that the corresponding LED is illuminated when each button or knob is selected. Figure 7-25 LEDs Tab...
  • Page 269 ANUAL 7-7-8 Ending the Program 1.) When you have finished testing the Op Panel, close the GE Test App. 2.) In the Service Platform window, click on the Stop button. 3.) Close the Service Platform window. 4.) Shutdown the system and reboot.
  • Page 270 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 7-8 Common Diagnostics Common Diagnostics is the second main heading under the GSUI heading of Diagnostics.
  • Page 271 USB Test • PCI Board Configuration Test • SCSI Device Test (This test normally fails since the Logiq 9 does not have any SCSI device) • AGP Video Card Test (This test is aborted, n/a for L9 BT04 or earlier) •...
  • Page 272 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 7-8-3 PC Diagnostics (Interactive Tests) • Audio Test (not available in BT06, BT07, or BT09) •...
  • Page 273 P R E L I M I N A R Y 1/18/10 CHAPTER7.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 7-9 Utilities 7-9-1 Dicom Verify Utility This utility provides an easy way to verify DICOM connectivity between the scanner and DICOM devices on the network.
  • Page 274 Whether the customer wants their data on CD or DVD media Tell the customer GE will attempt to recover as much data as possible. Write a complaint and include the above information, and do not discard or return the failed hard drive. A member of the Complaint Handling Unit or the Installed Base team will contact the author of the complaint and provide instructions for delivery of the failed hard drive to engineering.
  • Page 275 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems. Follow all safety precautions while servicing and repairing this equipment.
  • Page 276 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-1-2 Definitions of Left, Right, Front and Back Left Front Rear Right Figure 8-1 Console Views...
  • Page 277 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-1-3 Returning/Shipping Probes and Repair Parts Equipment being returned must be clean and free of blood and other infectious substances.
  • Page 278 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-2 Power Supply Battery Pack Replacement Procedure - BEP 1 8-2-1 Manpower One person, .5 hour + travel...
  • Page 279 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-2-5 Power Supply Battery Pack Installation Procedure - BEP 1 Heat-shrink holds batteries together In Line and insulates terminals.
  • Page 280 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-2-5 Power Supply Battery Pack Installation Procedure - BEP 1 (cont’d) The UPS batteries will drain slightly when the system is in the OFF condition. The OFF condition is:...
  • Page 281 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-3 Power Supply Battery Pack Replacement Procedure - BEP 2 8-3-1 Manpower One person, 15 minutes + travel...
  • Page 282 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-3-4 Power Supply Battery Pack Removal Procedure (cont’d) Figure 8-5 BEP 2 Power Supply Battery Pack Cover Plate Tipped Forward 4.) Disconnect the battery pack connection wire.
  • Page 283 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-3-5 Power Supply Battery Pack Installation Procedure - BEP 2 (cont’d) 8.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 284 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-4 Power Supply Battery Pack Replacement Procedure - BEP 3 / BEP 3.5...
  • Page 285 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-4-4 Power Supply Battery Pack Removal Procedure (cont’d) 3.) Disconnect the battery pack connection wires. See Figure 8-7.
  • Page 286 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-4-5 Power Supply Battery Pack Installation Procedure - BEP 3 / BEP 3.5 1.) Slide the new power supply battery packs into position in the Back End Processor.
  • Page 287 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-5 Power Supply Battery Pack Replacement Procedure - BEP 4.x 8-5-1 Manpower One person, 15 minutes + travel...
  • Page 288 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-5-4 Power Supply Battery Pack Removal Procedure - BEP 4.x (cont’d) 3.) Disconnect the battery pack connection wires.
  • Page 289 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-5-4 Power Supply Battery Pack Removal Procedure - BEP 4.x (cont’d) 5.) Take bracket off.
  • Page 290 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-5-5 Power Supply Battery Pack Installation Procedure - BEP 4.x 1.) Slide the new power supply battery packs into position in the Back End Processor.
  • Page 291 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-6 Before Starting a Software Load 8-6-1 Printers Any customized print settings will be lost when the new base image is loaded prior to R6.x.x. If the customer has any on board, peripheral, or network printers, any settings they have made in the device driver, such as Gray Scale Images, will be lost.
  • Page 292 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-7 Image Management Guide 8-7-1 Managing Patient Data and Presets Save to MOD, CD-R or DVD-R any patient data, images, and system presets that have been stored or configured.
  • Page 293 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-7-2 EZBackup/Move EZBackup/Move allows you to manage hard disk space (move images off the hard drive) while maintaining the patient database on the scanner, as well as to back up the patient database and images.
  • Page 294 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-7-2 EZBackup/Move (cont’d) Figure 8-16 New Patient EZBackup/Move Example NOTE: EZBackup/Move can take up to 20 minutes (or longer, depending on the size of the backup).
  • Page 295 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-7-2 EZBackup/Move (cont’d) NOTE: You do not need to format the media prior to performing an EZBackup/Move.
  • Page 296 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-7-3 User-Defined Backup Protocols 1.) Insert the backup media. Format the CD-R or MOD. (DVD-R do not need to be formatted.) Select the Utility tab on the Measurement Selection Menu.
  • Page 297 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-7-3 User-Defined Backup Protocols (cont’d) Figure 8-20 Format Successful Pop-up Menu 3.) On the Touch Panel, press Utility.
  • Page 298 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-7-3 User-Defined Backup Protocols (cont’d) Figure 8-21 Backup Menu- Prior to R4.x.x Figure 8-22 Backup / Restore Screen - R4.x.x and R6.x.x...
  • Page 299 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-7-3 User-Defined Backup Protocols (cont’d) NOTE: The detailed section of this menu decouples the user defined configuration above. This allows you to selectively restore what you want to restore across multiple machines.
  • Page 300 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-8 Loading Base Image Software - R4.x.x and prior 8-8-1 Preparations Save Connectivity Settings as described in 8-6-2 on page 8-17.
  • Page 301 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-8-1-1 To Prepare to Load Base Image (cont’d) 2.) Remove the left side cover.
  • Page 302 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-8-2 Base Image Load Procedure - R4.x.x and prior The next steps will destroy ALL data on the Scanner.
  • Page 303 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-8-2 Base Image Load Procedure - R4.x.x and prior (cont’d) 5.) Turn on the scanner. You will see a message displayed on the Touch Panel screen as shown in...
  • Page 304 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-8-2 Base Image Load Procedure - R4.x.x and prior (cont’d) Figure 8-27 Base Load Option Screen The first choice is to load the complete hard drive.
  • Page 305 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-8-2 Base Image Load Procedure - R4.x.x and prior (cont’d) The second choice is to load only the bootable portion of the hard drive partition, the “C” partition. This is not an option for a new hard drive that has not been previously partitioned for proper LOGIQ™...
  • Page 306 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-8-2 Base Image Load Procedure - R4.x.x and prior (cont’d) Once you have chosen to continue with Option 1 or Option 2, you will see a screen showing Norton Ghost.
  • Page 307 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-9 Loading Base Image Software - R6.x.x Contact the customer prior to this upgrade to allow them enough lead time to backup the system.
  • Page 308 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-9 Loading Base Image Software - R6.x.x (cont’d) 6.) Properly turn off the scanner by momentarily pressing the Standby (ON/OFF) Switch. Select Shutdown from the System Exit menu.
  • Page 309 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-9 Loading Base Image Software - R6.x.x (cont’d) Figure 8-33 Base Load Option Screen 10.)Select [1] Load the complete disk - All existing data will be lost.
  • Page 310 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-9 Loading Base Image Software - R6.x.x (cont’d) Once you have chosen to continue with Option 1, you see a screen showing Norton Ghost (see Figure 8-35.
  • Page 311 16.)You now need to run the Firmware executable to update the DVD/CD-ROM drives. a.) Determine the type of drive this LOGIQ 9 supports by right clicking on My Computer (left Set key) --> Manage (right Set key) --> Device Manager --> DVD/CD-ROM drives. Note the type of device.
  • Page 312 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-10 Loading Base Image Software - R7.x.x or Later Save all Service Presets if you are connected to InSite to restore remote connectivity after completed.
  • Page 313 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-10 Loading Base Image Software - R7.x.x or Later (cont’d) 8.) Properly turn off the scanner by momentarily pressing the Standby (ON/OFF) Switch. Select Shutdown from the System Exit menu.
  • Page 314 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-10 Loading Base Image Software - R7.x.x or Later (cont’d) Figure 8-40 Base Load Option Screen 12.)Select either option:...
  • Page 315 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-10 Loading Base Image Software - R7.x.x or Later (cont’d) Once you have chosen to continue with the selected option, you see a screen showing Norton Ghost...
  • Page 316 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-10 Loading Base Image Software - R7.x.x or Later (cont’d) 14.)Remove DVD from the CD-ROM/DVD drive.
  • Page 317 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-10 Loading Base Image Software - R7.x.x or Later (cont’d) Figure 8-46 DriveUpdater Dialog Box 18.)If the Current Version matches the New Version, proceed to...
  • Page 318 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-11 Loading Application Software - R4.x.x and prior 8-11-1 Loading Application Software - R4.x.x and prior NOTE: If you have not done so previously, double-click on the My Computer icon.
  • Page 319 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-11-1 Loading Application Software - R4.x.x and prior (cont’d) Figure 8-48 Command Window 6.) As indicated, press any key to continue.
  • Page 320 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-11-1 Loading Application Software - R4.x.x and prior (cont’d) Figure 8-50 BEP 3 Only - System Shutdown Warning Message After the system reboots, a DOS box displays showing you the installation of the Service Platform.
  • Page 321 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-11-1 Loading Application Software - R4.x.x and prior (cont’d) Figure 8-52 Start Software 10.)Remove the Applications CD.
  • Page 322 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-11-1 Loading Application Software - R4.x.x and prior (cont’d) 14.)Perform the following functional tests. If all are successful, include the debrief script provided below...
  • Page 323 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-12 Loading Application Software - R6.x.x 8-12-1 Loading Applications Software - R6.x.x 1.) Place the “Application Software” CD/DVD into the CDROM/DVD drive.
  • Page 324 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-12-1 Loading Applications Software - R6.x.x (cont’d) 5.) You see a window open similar to the example shown in Figure 8-48.
  • Page 325 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-12-1 Loading Applications Software - R6.x.x (cont’d) 7.) The system reboots before installing the Service Platform. You will see Figure 8-50 dialog briefly just before the system reboots.
  • Page 326 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-12-1 Loading Applications Software - R6.x.x (cont’d) 8.) When the software has completed loading, the system reboots.
  • Page 327 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-12-1 Loading Applications Software - R6.x.x (cont’d) If the Basic Option is not saved in the Option Dongle, then the SW License dialog window opens.
  • Page 328 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-12-2 Wired VoiceScan Speaker Assignment - R6.x.x, R7.x.x, and R9.x.x If the customer had the Wired VoiceScan option on their system, the following procedure needs to be done to change the audio output from the wired headset to the PC Speakers.
  • Page 329 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-12-2 Wired VoiceScan Speaker Assignment - R6.x.x, R7.x.x, and R9.x.x (cont’d) 3.) Select the Audio tab.
  • Page 330 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-13 Loading Application Software - R7.x.x or Later If you have previously performed a Base Image Software Load, proceed to section 8-23-1, Step 1.
  • Page 331 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-13 Loading Application Software - R7.x.x or Later (cont’d) 6.) A window similar to the example shown in Figure 8-65 opens.
  • Page 332 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-13 Loading Application Software - R7.x.x or Later (cont’d) 9.) The system reboots before installing the Service Platform. You will see...
  • Page 333 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-13 Loading Application Software - R7.x.x or Later (cont’d) 10.)When the software has completed loading, the system reboots.
  • Page 334 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-13 Loading Application Software - R7.x.x or Later (cont’d) NOTE: Starting with R3.x Software, all systems must have at least the basic option string in order for the software to run.
  • Page 335 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-14 Generic Application Software - Upgrade Only 8-14-1 Application Software Upgrade Procedure NOTE: When installing software you may see error messages about not being able to replace the following *.dll files: ginastub.dll and msvcr70.dll.
  • Page 336 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-14-1 Application Software Upgrade Procedure (cont’d) 4.) Select Install SW ..5.) Two separate dialog boxes open warning you that you are about to install new software as shown Figure 8-71.
  • Page 337 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-14-1 Application Software Upgrade Procedure (cont’d) 7.) As indicated, press any key to continue.
  • Page 338 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-14-1 Application Software Upgrade Procedure (cont’d) Figure 8-76 Enter Option String 12.)During the reboot process, the system may ask for the option string (valid license) to be entered.
  • Page 339 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-15 System Setup 8-15-1 To Complete the System Setup 1.) From the Touch Screen on the scanner, select Utility -> Connectivity -> TCPIP Tab.
  • Page 340 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-15-5 Print Key Assignment and Checkout Because of the addition of more printer choices, digital and analog, the factory default print key configuration has been removed for the color and b/w printers.
  • Page 341 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-15-6 Restore Patient, Report, and User-Defined Presets (cont’d) 3.) In the Restore list, select from the following, depending on which configurations you want to restore:...
  • Page 342 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-15-8 Standby Operation Checkout 8-15-8-1 To Place the System in the Standby Mode 1.) Press the ON/OFF switch at the front of the system once. The System-Exit window displays.
  • Page 343 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-15-9 Functional Checkout 1.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 344 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-16 Console Lock Assembly Replacement Procedure 8-16-1 Manpower One person, 1 hour 8-16-2...
  • Page 345 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-16-4 Console Lock Assembly Removal Procedure 1.) Unscrew 8 (5 mm) and 2 (3 mm) hexagon screws .
  • Page 346 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-17 Release Lock Assembly Replacement Procedure 8-17-1 Manpower One person, 1 hour 8-17-2 Tools •...
  • Page 347 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-17-5 Release Lock Installation Procedure 1.) Install Release Lock Assembly. 2.) Fasten 6 hexagon screws.
  • Page 348 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-18 Gas Shock Replacement Procedure 8-18-1 Manpower 2 persons, 4 hours Tilting the console requires two people in order to avoid injury to service personnel and...
  • Page 349 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-3 Overview (cont’d) You get access to the Gas Shock via openings in the frame behind the BEP, so the BEP must be removed.
  • Page 350 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-5 Top Tray Frame The Color Printer/Top Tray Frame should be remove to avoid damage to this part.
  • Page 351 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-5-1 Removal Procedure (cont’d) You should now have a chassis with all the necessary parts removed as shown in...
  • Page 352 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-6 Gas Shock Removal The stripped down system console will be tilted on its side in order to remove and replace the gas shock.
  • Page 353 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-6 Gas Shock Removal (cont’d) Bottom Shock Mount Top Shock Mount Figure 8-86 Console on its Side Beware of possible sharp edges on all metal chassis parts.
  • Page 354 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-6 Gas Shock Removal (cont’d) Release arm Wire end lock Spring Gas Spring Axle...
  • Page 355 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-7 Install the New Gas Shock Reverse the removal procedure to install the new gas shock.
  • Page 356 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-7 Install the New Gas Shock (cont’d) Mounting Pin “C” Clip Figure 8-91 Shock Bottom Mounted 2.) Place a washer on each side of the bottom of the shock mount and align the holes as shown in...
  • Page 357 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-7 Install the New Gas Shock (cont’d) Tilting the console requires two people in order to avoid injury to service personnel and...
  • Page 358 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-8 Reassemble Z-Release (cont’d) 3.) Thread the spring and barrel clamp(s) onto the Release Cable wire. Place the Barrel Clamp(s) in the Clamp cradle on the Shock’s Valve Release arm as shown in...
  • Page 359 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-18-8 Reassemble Z-Release (cont’d) 17.)Replace all covers. Air Filter Handle Top Cover Back Covers (Top & Bottom)
  • Page 360 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-19 Gas Spring Wirecable Replacement Procedure The Gas Spring Wirecable (FB307375) is a field replaceable unit. The wirecable is used when operating the Z-Release Handle to move the Top Console up/down.
  • Page 361 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-19-3 Wirecable Removal Preparations (cont’d) 5.) Remove VCR or DVR. Removing B/W Video Printer and VCR/DVR gives access to the lower end of wirecable and the release mechanism located behind and between the rear parts of these two units.
  • Page 362 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-19-4 Wirecable Removal Procedure 1.) Remove Z-handle shown in Figure 8-97. c.) Unscrew 1 hexagon screw.
  • Page 363 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-19-4 Wirecable Removal Procedure (cont’d) 4.) Pull the wirecable out of mantle (black rubber tube housing) from the upper end of the mantle shown Figure 8-99.
  • Page 364 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-19-5 Wirecable Installation Procedure 1.) Thread the new wirecable with prefixed wire-end lock through Z-Handle.
  • Page 365 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-19-5 Wirecable Installation Procedure (cont’d) 17.)Perform the following functional tests. If successful, include the debrief script provided below.
  • Page 366 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 8-20 Frog Leg Angle Adjustment 8-20-1 Manpower Two persons, .5 hour + travel...
  • Page 367 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-20-5 Frog Leg Adjustment Procedure (cont’d) 3.) Have one person hold on to the rear of the operator panel. See...
  • Page 368 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-20-5 Frog Leg Adjustment Procedure (cont’d) Figure 8-103 Remove Screws 6.) Have person holding the Operator Console gently tip it back. See...
  • Page 369 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-20-5 Frog Leg Adjustment Procedure (cont’d) 9.) Have person holding the Operator Console gently tip console back into place.
  • Page 370 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-20-6 Functional Check (cont’d) Thumbnail indicates 1/2 way down Crash Pad Figure 8-106 Crash Pad 3.) Lock and Unlock Frogleg to verify correct operation.
  • Page 371 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 8-20-6 Functional Check (cont’d) Table 8-8 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 372 P R E L I M I N A R Y 1/18/10 CHAPTER8.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 8-98 Section 8-20 - Frog Leg Angle Adjustment...
  • Page 373 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts available for the LOGIQ™...
  • Page 374 LOGIQ™ 9 220-50V PAL Console & Peripherals - BEP4.1 BT’07 + 5177000-2 BT09* LOGIQ™ 9 220-50V NTSC Console & Peripherals - BEP4.1 BT’07 + 5177000-3 BT09* *BT09 is a software-only upgrade Special Service Tool: service key/dongle (p/n 066E0703) required for some GE proprietary operations. Section 9-1 - Overview...
  • Page 375 P R E L I M I N A R Y 1/22/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-2 List of Abbreviations • Assy - Assembly • BEP - Back End Processor •...
  • Page 376 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-3 Console View Left Front Rear Right Figure 9-1 Console Views...
  • Page 377 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-4 External Covers 101A 102A 105A 103A 103B 104A 104B...
  • Page 378 Right Side Cover 103A FC200074 with bumper - Same as Vivid 7 - Right Side Cover Assembly with bumper and GE Logo. If ordering to replace 103B FC200422 non-logo cover, also order FC200421 and replace Right Side Cover Assembly FC307773...
  • Page 379 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-5 Additional Console Hardware 151A 145A 148A 150A 150B 147A...
  • Page 380 Left Probe Holder Assembly 149A 5212272 BT07 and later with 19” LCD only - Dark Steel Blue Left Probe Holder Assembly - DSB *Need to order GE Logo Decal 2269030-15 or 2291461 Gel Warmer Assembly * see note L9 Logo Decal 5255441 150A...
  • Page 381 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-6 Operator I/O Mechanical Parts 175A 182A 183A Figure 9-4 OP I/O Mechanical Parts...
  • Page 382 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-6 Operator I/O Mechanical Parts (cont’d) Table 9-6 Operator I/O Mechanical Parts...
  • Page 383 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-6 Operator I/O Mechanical Parts (Continued) Item Part Name Part Number...
  • Page 384 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-6-1 Bumper Kit, Frogleg Figure 9-5 Bumper Kit, Frogleg, Base Seen From Above...
  • Page 385 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-7 Operator Control Panel Brackets Bracket on units built before Mid-2002.
  • Page 386 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-8 Gas Spring WHEN THE TOP CONSOLE IS IN ITS LOCKED POSITION, THE GAS SPRING IS...
  • Page 387 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-9 Gas Spring (Continued) Item Part Name Part Number Description...
  • Page 388 Covers the screws on each side in the back of the 2303932 Monitor Screw Caps (covers) cover - Same as Vivid 7 & LOGIQ 7- 2303929-2 Includes three logos Vivid 7, LOGIQ 7 and LOGIQ 9 Monitor Front Bezel Assembly 2297050 User Switch Assembly Monitor Switch Assembly...
  • Page 389 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-9-2 Monitor Assembly - 17” LCD 261 261A 269A Figure 9-9 Monitor Assembly - LCD 17”...
  • Page 390 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-9-3 Monitor Assembly - 19” LCD 277A 286 286A 283A 281A...
  • Page 391 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-12 Monitor Assembly - 19” LCD (Continued) Item Part Name...
  • Page 392 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-10 Operator Control Panel Assembly 9-10-1 Upper Op Panel 201D...
  • Page 393 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-10 Operator Control Panel Assembly (cont’d) Table 9-14 Operator Control Panel Assembly - Upper...
  • Page 394 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-10-2 Lower Operator Panel 211A 211B 211C 211D 211E 212A 212B...
  • Page 395 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-15 Operator Control Panel Assembly - Lower (Continued) Item Part Name...
  • Page 396 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-10-3 Lower Op Panel - Buttons and Knobs 218 218A 219A...
  • Page 397 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-10-3 Lower Op Panel - Buttons and Knobs (cont’d) Table 9-16...
  • Page 398 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-11 Front End Processor Assembly 406A 402 402A BT’01 Card...
  • Page 399 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-11 Front End Processor Assembly (cont’d) Table 9-17 Front End Processor Assembly Applications Software R2.X.X...
  • Page 400 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-17 Front End Processor Assembly (Continued) Applications Software R2.X.X R3.X.X R4.X.X...
  • Page 401 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-17 Front End Processor Assembly (Continued) Applications Software R2.X.X R3.X.X R4.X.X...
  • Page 402 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-17 Front End Processor Assembly (Continued) Applications Software R2.X.X R3.X.X R4.X.X...
  • Page 403 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-17 Front End Processor Assembly (Continued) Applications Software R2.X.X R3.X.X R4.X.X...
  • Page 404 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-12 I/O Interfaces 302 302A 302B 302C 302D 302E USB MODEM OPTION...
  • Page 405 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-12 I/O Interfaces (cont’d) Table 9-18 I/O Interfaces Item Part Name...
  • Page 406 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-13 3-D Externals Figure 9-16 3-D Externals Table 9-19 3-D Externals...
  • Page 407 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-14 Back End Processor Assemblies 9-14-1 BEP 1 450 451...
  • Page 408 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-14-2 BEP 2 460A 460A 460A 460B 460C 460C 460C 460C...
  • Page 409 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-21 Back End Processor Assembly 2 (Continued) Item Part Name...
  • Page 410 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-14-3 BEP 3.x / 4.x 475A 475B 481 481A 475C 478A BEP3.x...
  • Page 411 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-14-3BEP 3.x / 4.x (cont’d) Table 9-22 Back End Processor Assembly 3.x / 4.x...
  • Page 412 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-22 Back End Processor Assembly 3.x / 4.x (Continued) Item...
  • Page 413 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-15 Software Chapter 9 9-41...
  • Page 414 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-15 Software (cont’d) 9-42 Section 9-15 - Software...
  • Page 415 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-15 Software (cont’d) Table 9-25 Software Tools Item Part Name...
  • Page 416 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-16 Interconnect Cables 9-16-1 Peripheral Cables TO CARD RACK SONY VCR SVO-9500MD...
  • Page 417 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-1 Peripheral Cables (cont’d) Table 9-26 Peripheral Cables Item Part Name...
  • Page 418 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-26 Peripheral Cables (Continued) Item Part Name Part Number Description...
  • Page 419 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-26 Peripheral Cables (Continued) Item Part Name Part Number Description...
  • Page 420 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-2 Peripheral Cable - Analog Color Printer with BEP3 TO CARD RACK...
  • Page 421 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-3 Peripheral Cable - USB Printer Cables BW PRINTER UP-D895 PRINTER...
  • Page 422 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-4 AC Distribution Cables To Color Printer To Wireless Mic To Color Printer...
  • Page 423 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-4 AC Distribution Cables (cont’d) Table 9-29 AC Distribution Cables Part...
  • Page 424 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-29 AC Distribution Cables (Continued) Part Item Part Name Number...
  • Page 425 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-29 AC Distribution Cables (Continued) Part Item Part Name Number...
  • Page 426 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-4 AC Distribution Cables (cont’d) Figure 9-24 BEP 1.0 and II/O Cabling Diagram...
  • Page 427 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-4 AC Distribution Cables (cont’d) Figure 9-25 BEP 2.0 and II/O Cabling Diagram...
  • Page 428 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-4 AC Distribution Cables (cont’d) Figure 9-26 BEP 2.1/2.2 and II/O Cabling Diagram...
  • Page 429 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-4 AC Distribution Cables (cont’d) Figure 9-27 BEP 3.x/BEP 4.x and II/O Cabling Diagram...
  • Page 430 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-16-4 AC Distribution Cables (cont’d) Table 9-30 Internal I/O Cables Part...
  • Page 431 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-30 Internal I/O Cables (Continued) Part Item Part Name Number...
  • Page 432 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 9-30 Internal I/O Cables (Continued) Part Item Part Name Number...
  • Page 433 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-17 Power Distribution Modules 608A 604A 603A Figure 9-28 Power Distribution...
  • Page 434 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-18 Air Distribution Figure 9-29 Air Distribution Parts Table 9-32...
  • Page 435 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-19 Voice Scan 750 756 757 764 765A 761A 767A...
  • Page 436 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-19 Voice Scan (cont’d) Table 9-33 Voice Scan Item Part Name...
  • Page 437 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 9-19-1 Voice Scan Cabling WIRELESS MIC OPTION Distribution 2383769 2384314 NOTE:...
  • Page 438 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-20 4D Components 800A 803 804 Figure 9-32 4D Components...
  • Page 439 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-21 Continuous Wave (CW) Doppler Components Figure 9-33 Continuous Wave (CW) Doppler Components...
  • Page 440 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-22 Peripherals Table 9-37 Approved Peripherals Part Name Part Number...
  • Page 441 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-23 Customer Documentation Table 9-38 User Documentation BT02 BT03 BT04...
  • Page 442 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-24 Probes Figure 9-34 LOGIQ™ 9 Probes 9-70 Section 9-24 - Probes...
  • Page 443 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-24 Probes (cont’d) Table 9-39 Available Probes Item Part Name...
  • Page 444 P R E L I M I N A R Y 1/18/10 CHAPTER9.FM GE H PROPRIETARY TO GE EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 9-24 Probes (cont’d) Table 9-40 Software/Probe Compatibility Matrix Item Part Name Part Number 7.x/9.x...
  • Page 445 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your LOGIQ™ 9 system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
  • Page 446 Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
  • Page 447 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-3-1 How often should care & maintenance tasks be performed? (cont’d) Table 10-2 Customer Care Schedule...
  • Page 448 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Care & Maintenance Table 10-3 Overview of Requirements for Care &...
  • Page 449 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary.
  • Page 450 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary.
  • Page 451 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
  • Page 452 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-5-4 Cleaning 10-5-4-1 General Cleaning Table 10-8 General Cleaning Step Item Description Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful Console not to get the cloth too wet so that moisture does not enter the console.
  • Page 453 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-5-5 Physical Inspection Table 10-10 Physical Checks Step Item Description Verify that all system labeling is present and in readable condition. Refer to the LOGIQ™ 9 User Labeling Manual for details.
  • Page 454 The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
  • Page 455 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 10-7 Electrical Safety Tests 10-7-1 Safety Test Overview The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards.
  • Page 456 ANUAL 10-7-2 GE Healthcare Leakage Current Limits for LOGIQ™ 9 The following limits are summarized for NFPA 99 (For USA), IEC 60601-1 Medical Equipment Safety Standards, and IEC 62353 Medical Electrical Equipment — Recurrent test and test after repair of medical electrical equipment.
  • Page 457 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 458 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-4 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test CAUTION Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
  • Page 459 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break.
  • Page 460 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-5-2 Generic Procedure The test verifies the isolation of the power line from the chassis. The testing meter is connected from accessible metal parts of the case to ground.
  • Page 461 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-5-3 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-12.
  • Page 462 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On.
  • Page 463 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-7 Isolated Patient Lead (Source) Leakage–Lead to Lead Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function selector to the LEAD-LEAD position.
  • Page 464 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the Planned Maintenance Inspection Certificate.
  • Page 465 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-9 Probe Leakage Current Test 10-7-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 466 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-9-2 Test Equipment (cont’d) Table 10-23 Test Equipment and Accessory Description Dale Part number...
  • Page 467 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-9-3 Generic procedure on Leakage Current Measurements should be made with the ground open and closed, with power line mains polarity normal, and with the unit Off and On.
  • Page 468 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-9-4 Meter procedure using probe adapter (cont’d) Equipment damage possibility. Never switch the Polarity or the status of the Neutral when the CAUTION Ultrasound unit is powered on.
  • Page 469 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-9-5 Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins.
  • Page 470 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-7-9-5Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current (cont’d) 5.) Set the meter's “FUNCTION” switch to EXTERNAL (DALE 600) or Mains on Applied Parts (M.A.P.) (DALE 601) position.
  • Page 471 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 10-8 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
  • Page 472 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL 10-8-1 ULTRASOUND INSPECTION CERTIFICATE ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.)
  • Page 473 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Section 10-9 Site Log Table 10-26 Site Log Date Service person Problem Comments Chapter 10 Care & Maintenance...
  • Page 474 P R E L I M I N A R Y 1/18/10 CHAPTER10.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Table 10-26 Site Log (Continued) Date Service person Problem Comments 10-30 Section 10-9 - Site Log...
  • Page 475 P R E L I M I N A R Y 1/18/10 L9_IX.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL INDEX Numerics Brake Adjustment, 6-15 3-D Externals, 9-34 Brake Arm Adjustment, 6-13 Brake Pedal...
  • Page 476 P R E L I M I N A R Y 1/18/10 L9_IX.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL EQ Board Basic Functions, 5-17 I/O Ports, 3-17 General Description, 5-16 Image Management Equalization (EQ) Board, 5-16...
  • Page 477 P R E L I M I N A R Y 1/18/10 L9_IX.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Modem, 5-69 Probes/Connector Usage Inputs/Outputs, 5-69 Functional Checks, 4-11 LEDs, 5-70 Product Icons, 1-11...
  • Page 478 P R E L I M I N A R Y 1/18/10 L9_IX.FM GE H EALTHCARE 2294853-100, R LOGIQ™ 9 B IRECTION EVISION ASIC ERVICE ANUAL Touch Panel Functional Checks, 4-3, , 4-4 Trouble Image with Logs, 7-4 Troubleshooting Gathering Trouble Data, 7-3...
  • Page 480 , General Electric Company. © 2002-2010 GE Healthcare - GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC. 9900 Innovation Drive Wauwatosa, Wisconsin 53226 www.gehealthcare.com...