Installation; Safety; System Design Considerations - Edwards RV3 Instruction Manual

Rotary vane pumps
Hide thumbs Also See for RV3:
Table of Contents

Advertisement

3

Installation

3.1

Safety

We recommend that you do not use a hydrocarbon-prepared RV pump for pumping hazardous
substances. PFPE-prepared pumps are suitable for oxygen applications: refer to
You must ensure that the RV pump is suitable for your application. If you have any doubt as to the suitability of the
RV pump for your application, refer to the Edwards guidelines on vacuum pump and vacuum system safety (see the
Associated publications at the end of the Contents list at the front of this manual).
A suitably trained and supervised technician must install your RV pump. Obey the safety instructions listed below
when you install the pump, especially when you connect the pump into an existing system. Details of specific safety
precautions are given at the appropriate point in the instructions.
Wear the appropriate safety-clothing when you come into contact with contaminated components.
Vent and purge your vacuum system before you start installation work.
Ensure that the installation technician is familiar with the safety procedures which relate to the pump-oil
and the products handled by the pumping system. Take suitable precautions to avoid the inhalation of oil
mist and excessive skin contact with pump-oil, as prolonged exposure can be harmful.
Disconnect the other components in the pumping system from the electrical supply so that they cannot be
operated accidentally.
3.2

System design considerations

Consider the following points when you design your pumping system:
Use a suitable valve to isolate the pump from your vacuum system if you need to allow the pump to warm up
before you pump condensable vapours, or to provide additional system protection when the pump is
switched off.
Avoid high levels of heat input to the pump from the process gases, otherwise the pump may overheat and
seize, and cause the motor thermal overload device to open.
If you use the pump in a high ambient temperature and have a high gas throughput, the temperature of the
pump-body may exceed 70°C and you must fit suitable guards to prevent contact with hot surfaces.
Make sure that the exhaust pipeline cannot become blocked. If you have an exhaust-isolation valve, make
sure that you cannot operate the pump with the valve closed.
Provide for a purge of inert gas when you shut down the pumping system, to dilute dangerous gases to safe
concentrations. A suitable gas ballast adaptor for introduction of purge gas into the pump is available as an
accessory (see
Section
© Edwards Limited 2007. All rights reserved.
Edwards and the Edwards logo are trade marks of Edwards Limited.
WARNING
7.4.8).
A652–01–880 Issue S
Section
8.
Page 17