5

Maintenance

5.1

Safety information

Obey the safety instructions given below and take note of appropriate precautions. If you do not,
you can cause injury to people and damage to equipment.
If your pump is PFPE-prepared, refer to
A suitably trained and supervised technician must maintain the pump. Obey your local and national safety
requirements.
Ensure that the maintenance technician is familiar with the safety procedures which relate to the pump-oil
and the products processed by the pumping system.
Check that all the required parts are available and of the correct type before you start work.
Isolate the pump and other components from the electrical supply so that they cannot be operated
accidentally.
Allow the pump to cool (so that it is at a safe temperature for skin contact) before you start maintenance
work. Make sure the pump is switched off in case the thermal overload device restarts the pump.
Do not reuse 'O' rings and seals if they are damaged.
After maintenance is completed, recheck the direction of pump rotation if the electrical supply has been
disconnected.
The pump and the pump-oil will be contaminated with the process chemicals that have been pumped during
operation. Ensure that the pump is decontaminated before maintenance and that you take adequate
precautions to protect people from the effects of dangerous substances if contamination has occurred.
Do not touch or inhale the thermal breakdown products of fluorinated materials which may be present if the
pump has been heated to 310°C and above. Fluorinated materials are safe in normal use but can decompose
into very dangerous substances (which may include hydrofluoric acid) if they are heated to 310°C and above.
The pump may have overheated if it was misused, if it malfunctioned, or if it was in a fire. Material Safety
Data Sheets for fluorinated materials used in the pump are available on request: contact your supplier or
Edwards.
If necessary, maintain the motor as specified in the manufacturers information supplied with the motor.
© Edwards Limited 2007. All rights reserved.
Edwards and the Edwards logo are trade marks of Edwards Limited.
WARNING
Section 8
before you maintain the pump.
A652–01–880 Issue S
Page 33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rv5Rv8Rv12

Table of Contents