Storage and disposal

6.1

Storage

Observe the storage temperature limits stated in
pump seals.
Note:
If you will store a new pump in conditions of high humidity, remove the pump from its cardboard packaging
box; dispose of the box (refer to
Use the following procedure to store the pump:
1. Shut-down the pump as described in
2. Disconnect the pump from the electrical supply.
3. Purge your vacuum system and the pump with dry nitrogen and disconnect the pump from your vacuum system.
4. Replace the oil as described in
5. Place and secure protective covers over the inlet and outlet-ports.
6. Store the pump in cool, dry conditions until required for use. When required, prepare and install the pump as
described in
Section
3. If the pump has been stored for more than a year, before you install the pump you must
clean and overhaul it as described in the instructions supplied with the clean and overhaul kit.
6.2

Disposal

Dispose of the pump and any components removed from it safely in accordance with all local and national safety and
environmental requirements.
Take particular care with components and waste oil which have been contaminated with dangerous process
substances.
Do not incinerate fluoroelastomer seals and 'O' rings.
© Edwards Limited 2007. All rights reserved.
Edwards and the Edwards logo are trade marks of Edwards Limited.
CAUTION
Section
2.1. Storage below -30°C will permanently damage the
Section
6.2).
Section
4.8.
Section
5.4.
A652–01–880 Issue S
Page 41