Daily Start Up

3.8 Start Analysis

7
8
9
10
3.8 Start Analysis
The reaction time is approximately 8 minutes and 40 seconds for the first result
to be obtained after the sample is dispensed. Every 3.6 seconds two tests can be
sampled. Each analyzer unit has a maximum throughput of 2,000 tests per
hour.
Results can be viewed on the monitor and printed.
To start analysis:
1
3-32
Select
to register the QC requisition.
Entry (F2)
Select
Display Cup Set (F5)
positions required for the control(s). Load the controls according to
the list in the correct green racks. Select
TIP In Display Cup Set (F5), the Volume (µL) is the sample volume
required based on the requisitioned tests. The dead volume is
not included.
If QC barcode operation is in use, the Rack No. in Display Cup
Set (F5) displays as 000x.
Load the racks on the rack set trays on the rack feeder unit. Load the
green rack(s) in numerical rack ID order.
Select
Start
.
Select the
Start
button from the main button bar. The Start window
displays with an Error List if any errors are present. Review errors
carefully and perform necessary corrective actions before starting
analysis. If an error is in red, it will be necessary to perform corrective
actions before that unit(s) can be started.
to view the control(s), racks, and
Close
to close the window.
B00418AA