Daily Start Up
3.5 Check Reagent Status
Item
Type:
R1/R2:
Shots:
Onboard Remaining:
RB Stability Remaining:
Cal Stability Remaining:
Expiration:
Bottle No. (SN):
Seq.:
Comment:
Edit (F1)
Position Setting (F2)
ID Edit (F3)
blue "*":
TIP The time remaining displays in hours (H) up to 72 hours, and days (D) over
72 hours.
If 2 tests are programmed to use one reagent, the RB Stability Remaining
and Cal Stability Remaining display as lower Test No./higher Test No.

3.5.3 Reagent Replacement

If the reagent volume is insufficient, on-board stability has expired, or the
reagent is expired, remove the old reagent bottles and replace them with a new
set.
CAUTION
Never add new reagent to existing reagent in Beckman Coulter
reagent ID bottles.
3-18
Select Serum, Urine, Other-1, Other-2
R1-1 and R2-1 are standard. R1-2 is for a test with a 3-part reagent.
Number of tests remaining in the bottle
Hours (H) or Days (D) remaining until the reagent on-board stability expires.
Hours (H) or Days (D) remaining until the reagent blank stability expires.
Hours (H) or Days (D) remaining until the calibration stability expire.
The expiration date of the reagent lot number.
A unique 4-digit number to identify each bottle of reagent.
Sequence number 1-5 of the same reagent in the R1/R2.
A caution or error message for the reagent.
Edit the test name, lot number, bottle number, and bottle size for fixed reagents.
Assign a position as Reagent ID or Fixed. When the position is fixed, an * appears
to the left of the Position number.
Edit the 20-digit reagent ID. Use this option to edit the reagent ID after a reagent ID
read error.
A fixed reagent position.
Description
B00418AA