6.2 Create a Profile

A profile is a group of tests that are usually requested at the same time. Using a
profile reduces the number of key strokes needed, as a single profile is selected
instead of multiple tests. A maximum of 100 profiles (Number 0 to Number
99) can be registered for samples, reagent blank, calibration, and QC. A
maximum of 120 tests can be programmed in a profile. The number of tests
that can be programmed in a profile is limited by the number of analyzer units,
sample blank tests, LIH, and sample type.
Each profile is assigned a profile name.
TIP Profile 0 is the Default profile in the Sample tab. Profile 0 is automatically
performed when:
TIP Tests that are grayed out cannot be selected.
ISE tests can be selected only when the sample type is "Serum" or "Urine".
B00418AA
Run the test to verify the programming of the new test(s).
• Load the reagent and any required cleaning solution on the
analyzer.
• Perform a reagent volume check.
• Verify that calibration is requisitioned for the new test(s).
• Requisition QC on the new test(s) if the test(s) was not added to
the default QC requisition.
• Perform a reagent blank, calibration, and QC on the new test(s).
• Review the printout and verify the reagent blank, calibration, and
QC data.
There is a barcode read error.
There is no requisition found for a sample.
There are online errors.
As Needed Tasks
6
6.2 Create a Profile
6-9