Program Calibrator Concentrations - Beckman Coulter AU5800 Quick Manual

Chemistry analyzer
Hide thumbs Also See for AU5800:
Table of Contents

Advertisement

6.3 Program Calibrator Concentrations

Use this function to verify and change calibrator concentrations. Select the
calibrator name to verify and change the concentrations of all tests in a
calibrator from the same screen. This is the easiest way to change all
concentrations when the calibrator lot number changes.
CAUTION
Confirm calibrator concentration values in Calibration Specific.
IT IS CRITICAL ALL MULTI-POINT CALIBRATOR VALUES WERE
ENTERED CORRECTLY.
For details on calibration specific test parameters, refer to 4.5.3 Calibration
Specific Menu in the AU5800 User's Guide, chapter 4.
1
2
3
B00418AA
Select
Menu List > Parameters > Calibration Parameters >
.
Calibrators
Select
. The window changes to the edit mode.
Edit (F1)
Select the calibrator "Name" to edit.
As Needed Tasks
6.3 Program Calibrator Concentrations
6
6-15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents