Fix (Assign) A Reagent - Beckman Coulter AU5800 Quick Manual

Chemistry analyzer
Hide thumbs Also See for AU5800:
Table of Contents

Advertisement

As Needed Tasks
6.9 Fix (Assign) a Reagent Position
1
2
3
6.9 Fix (Assign) a Reagent Position
Fix (assign) reagents that are not barcoded. The analyzer does not read
barcodes on reagents in fixed positions.
Non-barcoded reagents must be placed in the assigned reagent position.
Reagent ID Read Errors only display in the
Display" is set to Position.
1
2
3
4
6-30
From the
RB/CAL/QC
transfer.
Select
Transfer to Online (F7)
window.
Select
. Online transfer is performed.
OK
TIP Select Online Transfer Stop (F7) to stop the transfer.
Select
Home > Reagent Management > Details
TIP Reagent Status may have to be "checked" before a position can
be fixed. Select Reagent Check (F5) > "Reset".
Select the Unit No.
Select
from the "Reagent Display" drop-down list.
Position
Select an available bottle position to fix the reagent.
screen, select the samples and criteria to
to display the Transfer to Online
tab when "Reagent