Program A New Test - Beckman Coulter AU5800 Quick Manual

Chemistry analyzer
Hide thumbs Also See for AU5800:
Table of Contents

Advertisement

As Needed Tasks

6.1 Program a New Test

6.12 Reagent Consumption
6.13 Edit Quality Control Data
6.14 Save Data to External Media
6.15 Save or Load Parameters
6.16 Disable a Test
6.1 Program a New Test
Program a new test using the current parameter sheet. Refer to the AU5800
User's Guide, CHAPTER 4, Parameters for detailed Parameter menu
descriptions.
1
CAUTION
Changing the test name affects all results associated with that test
number. Any previously reported results (with the old test name) are
assigned the new test name. Extreme caution must be applied when
making any changes to the test name.
DO NOT CHANGE THE TEST NAME WITHOUT NOTING THE TIME
AND DATE THAT THE CHANGE OCCURRED AND THEN ENSURING
ANY RESULTS PRINTED OUT PRIOR TO THIS TIME AND DATE ARE
REVIEWED AND CORRECTLY IDENTIFIED.
6-2
6.12.1 Check Reagent Consumption by Reagent Dispense
6.12.2 Check Reagent Consumption by Samples Measured
6.15.1 Saving or Loading Parameter Files
Select
Menu List > Parameters > Common Test Parameters
display the
Test Name
• Select
.
Edit (F1)
• Test Name: Enter a maximum of 6 characters.
tab.
to
B00418AA