Edit A Reagent Id - Beckman Coulter AU5800 Quick Manual

Chemistry analyzer
Hide thumbs Also See for AU5800:
Table of Contents

Advertisement

As Needed Tasks

6.10 Edit a Reagent ID

9
10
6.10 Edit a Reagent ID
Edit the reagent ID after a reagent ID read error occurs on a barcoded reagent
bottle.
A notification alarm occurs during the reagent check ("Reagent ID Read
Error"), and the Comment "ID Edit" displays until the reagent bottle is
removed from the refrigerator.
1
2
3
4
5
6-32
Verify a blue box with an "*" displays to the left of "Pos." for fixed
reagents.
Select
Reagent Check (F5) > Check Specified Positions
the shot count.
Select
Home > Reagent Management > Details
Select the Unit No.
Select
from the "Reagent Display" drop-down list.
Position
Place the cursor on the position with the reagent ID read error.
Select
.
ID Edit (F3)
to update
.
B00418AA