Daily Start Up

5.8 Start Analysis

5.8 Start Analysis
The reaction time is approximately 8 minutes and 40 seconds for the first result to be obtained after
the sample is dispensed. Every 3.6 seconds, two tests can be sampled. Each analyzer unit has a
maximum throughput of 2,000 tests per hour.
Results can be viewed on the monitor and printed.
To start analysis:
1
Select
Start
Figure 5.4
Review errors carefully and perform necessary corrective actions before starting analysis. If an
error is in red, it is necessary to perform corrective actions before that unit(s) can be started.
2
Verify all units have a check in the box indicating the unit will be started. If the box is gray and
inaccessible, review the error list and perform necessary corrective actions before starting.
Remove the check from the box if analysis is not required on a unit, and the unit will not start.
3
Select
Start
changes from Standby to Measure 1.
5-6
to display the Start window with an Error List if any errors are present.
. The system initializes, and analysis starts if no errors are detected. The mode
PN B08501AB