Advertisement

Monitoring Results

5.1 Monitor Results

5.1 Monitor Results
It is important to verify daily reagent blank, calibration, and control (QC)
results are acceptable before reporting sample results. All results including
reagent blank, calibration, QC, and samples should be reviewed for error flags
and appropriate corrective actions taken before reporting the results. In
addition, the
taken for any alarms generated.
Refer to 7.4 Check Results in the AU5800 User's Guide, chapter 7, for more
information on monitoring reagent blank, calibration, QC, and sample results.
For more information on troubleshooting and corrective actions, refer to 11.3
Troubleshooting the System - Reagents, Calibrators, Quality Control and
Samples in the AU5800 User's Guide, chapter 11.

5.2 Identifying Sample Types by Sample Data Prefix

Sample Data Prefix
5-2
should be reviewed and appropriate corrective actions
Alarm List
Sample
R
A
Q
None
U
X
Y
R001
Reagent Blank
A001
Calibrator
Q001
Quality Control (QC)
0001
Routine Serum
U001
Routine Urine
X001
Routine Other-1
Y001
Routine Other-2
Sample Type
B00418AA