6.12.1 Check Reagent Consumption by Reagent Dispense
1
2
B00418AA
Shot Total
TIP The
analyzer from the installation of the instrument.
Select
Serum
,
Urine
list.
Select
Display Range (F3)
Select the start index and the end index from each drop-down list.
Select
OK
. The number of cumulative tests is displayed.
Select
after setting the index range and sample type. The
Test Shots
number of reagent dispenses for each test and sample type displays.
or
Select
R1 Volume
The volume in mL of reagent dispensed for each test and sample type
displays.
The reagent consumption is calculated by the following formula:
Actual analysis result * (the amount of reagent dispensing + the
amount of surplus dispensing)
is the cumulative number of tests run on the
,
Other-1
or
Other-2
.
.
R2 Volume
As Needed Tasks
6.12 Reagent Consumption
from the "Type" drop-down
6
6-37