Processing Samples

4.3 Place a Set Tray on the Analyzer

4.3 Place a Set Tray on the Analyzer
Two trays can be placed on the rack set unit, and two trays can be placed on the
rack collection unit.

4.3.1 Place Trays

1
WARNING
Before removing the tray from the system during operation, confirm
that the orange LED is off. If the tray is removed while the orange
LED is lit, an injury may result or the rack may topple over.
4-10
Confirm that the orange LED on the rack set unit and rack collection
unit are not blinking.
Figure 4.4
For details of LED functions, refer to 3.4.3 LEDs and Operation
Buttons (Rack Feeder Unit) in the AU5800 User's Guide, chapter 3.
1. Rack collection unit Orange LED
2. Rack set unit 1 Orange LED
3. Rack set unit 2 Orange LED
B00418AA