As Needed Tasks
6.13 Edit Quality Control Data

6.12.2 Check Reagent Consumption by Samples Measured

1
6.13 Edit Quality Control Data
Analyzed QC data can be searched and edited.
Editing can be performed by QC sample number or test.
CAUTION
Analyzed QC results can be edited. To prevent an erroneous
diagnosis due to numerous changes to the quality control data,
editing should be done according to laboratory protocol.
TIP When the QC analysis data results have been edited, confirm that the
edited data falls within the cumulative period. If it falls within the cumulative
period, the editing of the contents must be reflected in the cumulative
values. To reflect the editing of contents in the cumulative values, update
the cumulative values. Refer to 4.6.2 QC Specific Menu in the AU5800
User's Guide, chapter4 for updating and calculating cumulative QC values.
TIP When test results have been edited, an "e" flag is added to the Data Flags.
Once an "e" flag has been added, it cannot be deleted. When a "d" flag is
added to the Data Flags, the data is not included in the QC statistic
calculations.
6-38
Select
after setting the index range and sample type. The
Test Total
number of samples measured for each test and sample type displays.
B00418AA