Sirona Dental Systems GmbH
Service Manual INTEGO / INTEGO pro
X15
LED
Color
36 V
Green
5 V
Green
26.5V
Green
RUN
Green
Error
Red
Reset
Red
BOOT
yellow
S3
Red
AMALG
yellow
MOTOR
yellow
SUCTION
yellow
64 57 316 D3543
D3543.076.01.04.02 06.2015
6.4.2

DWEm board compact water unit control

X13 X12
X8
DS30
Components on the board
LEDs
Function
36 V input voltage
generated +5 V voltage
Generated +26.5 V voltage
Flashes 1x per second during normal operation
Illuminated: Error involving solenoid valves
2x flashes: Loss of communication with DSA
Lights up while the board is performing a reset
Lights up after switching on and remains lit until completion of the boot process
Illuminated: Amalgam separator tank disassembled
Illuminated: Amalgam separator 95%
Flashing: 100% full
Illuminated: Amalgam motor active
Illuminated: Suction machine active
X1
X45
X31
X30
X4
X7
S3
6 Overview of modules and boards
6.4 Water unit
X21
X2
X44
CAN-
CAN+
36V
GND