Care and Maintenance

Care and Maintenance
The advanced digital circuitry within the CO-Pilot system components requires
no calibration or periodic maintenance.
Do not attempt to open the case of any of the system components or repair the
electronics. Opening the case will damage the component and void the
warranty. If the device or system is not functioning properly, see
"Troubleshooting" section on page 42.
The Oxitest Plus 7 (software rev. 2.5 or greater) by Datrend Systems, Inc. can
be used to verify operation of the pulse oximeter. The Oxitest cannot be used to
verify COHb, MetHb, or rSO
CAUTIONS:
!
• Follow local, state and national governing ordinances and
recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
• Batteries are a fire hazard if damaged. Do not damage,
mishandle, disassemble, service, or replace with non-specified
components.
• Do not charge Li-Ion batteries at a temperature of 0 °C (32 °F) or
less as this may result in significantly reduced battery life.

Cleaning Instructions

The following cleaning instructions apply to the display and signal processor.
1.
Wipe the component with a soft cloth dampened with a cleaning solution,
taking care to avoid liquid entering open connectors (leave reusable
sensors connected to the signal processor; ensure display charging port
cover is fully inserted; do not use excessive liquid). Do not use any
cleaning solution other than those listed below, as permanent damage
could result.
2.
Allow to air-dry.
Approved cleaning chemicals:
• Isopropyl Alcohol
• ammonium chloride
• 10% bleach/90% water solution
The following cleaning instructions apply to the carrying case.