Device Pairing - Nonin Avant 4100 Operator's Manual

Digital pulse oximetry system with bluetooth wireless technology
Table of Contents

Advertisement

Using the Avant™ 4000 Digital Pulse Oximetry System

Device Pairing

Use the following procedure to pair the display unit with a desired wrist-worn
patient module.
NOTE: To ensure proper pairing, disconnect and then reconnect the
sensor from the patient module before beginning.
NOTE: The Avant 4000 display unit must be OFF before pairing devices.
1.
Press and hold the Bluetooth Control button while turning on the display unit
to begin the device pairing process. Notice that amber dashes flash in the
SpO
and Pulse Rate display areas as the system searches for patient modules
2
available for pairing. (If a patient module is available, its identification number
will appear on the display unit screen within 10 seconds.)
2.
When available patient module(s) are detected, a maximum of five "pairable"
patient modules will be displayed separately, with the device identification
number appearing in the SpO
identification number is included on the sides of each patient module.)
3.
If multiple patient modules are available for pairing, use the Plus or Minus
buttons to scroll through the available modules. (While five devices may be
available, only one may be selected for pairing.)
4.
When the identification number of the patient module you wish to pair with is
shown—and is not flashing—press the Bluetooth Control button again to
complete the pairing process. Normal operation resumes automatically when
the pairing process is complete.
5.
To confirm that the pairing process was successful, ensure that the
Connection Status LED is lit green.
NOTE: After the display unit is paired to a patient module, it will remain
paired until the above process is repeated.