Modes

The WristOx
Model 3150, has two types of mode settings: Activation and Display, which are
2,
described below.

Activation Modes

Activation modes determine how the 3150 turns off and on. Spot Check Activation Mode is the
factory default.

Spot Check Activation Mode

In this activation mode, the 3150 will automatically turn on when a finger is inserted into the
sensor. It will turn off 10 seconds after the finger is removed. If the sensor is disconnected, the
device turns off immediately.
NOTE: If the device determines that a sensor fault exists (a sensor failure, misalignment,
or incompatibility with the device) or if a pulse oximeter sensor signal cannot be detected,
the device turns off after 3 minutes.

Sensor Activation Mode

In this mode, the device turns on when the activation switch is pressed or when the sensor is
disconnected and reconnected. This mode is useful when using a sensor that is not easily
removed from the patient (e.g., disposable or wrap sensor).
If the sensor is not used or an inadequate pulse signal is detected for at least 10 minutes, the
device will turn off. To turn the device on again, press the activation switch or disconnect and
reconnect the sensor.

Programmed Activation Mode

When in Programmed Activation Mode, the user can program the device to turn on for up to
three sessions. In between the programmed sessions, the device will turn off. When off, the
device will display the next session start time every 30 seconds.
CAUTION: When setting the clock in Programmed Activation Mode using nVISION
!
software, verify all session dates and times are valid.
A sensor must be connected to the 3150 in order for the 3150 to turn on at the programmed
session time.
Between the programmed sessions, pressing the activation switch will turn on the device for
three minutes. During this time, the user is able to take and store measurements. After three
minutes, the device will turn off.
CAUTION: A device in Programmed Activation Mode reverts to Spot Check
!
Activation Mode if the clock is not set or if the clock settings are lost when replacing
the batteries or due to critical battery level.