Event Marker; Carbon Dioxide Detector Startup; Setup; Setup Mode - Nonin 9847V Operator's Manual

Veterinary handheld pulse; oximeter and carbon dioxide (co2) detector
Table of Contents

Advertisement

Event Marker

Press Alarm Limit Review to activate an event marker. Marked events will display as an asterisk (*)
at the end of the real-time data output.

Carbon Dioxide Detector Startup

Verify:
1. Airway adapter tube is properly attached to the CO
2. CO
sensor is properly connected to the device.
2
3. Lower bar of the CO
4. CO
sensor light is blinking.
2
5. CO
bar graph segments indicate change in CO
2
Attach the CO
sensor/airway adapter tube to the animal's endotracheal tube. The CO
2
reflect values and breath beeps for each breath (if the breath beep sound volume is not turned off).
A medium priority (equipment) alarm will occur (if the alarms are not disabled), when:
• The CO
sensor is disconnected.
2
• The airway adapter tube is removed from the CO
• The light path is blocked.
• A CO
sensor failure occurs.
2
High priority (animal) alarms override medium priority (equipment) alarms. The third and sixth
segments of the CO
bar graph will be lit, indicating a CO
2

Setup

All functions of the 9847V are controlled by buttons found on the
keypad on the front of the device.
Press On/Standby to turn the device on or to enter Standby mode.

Setup Mode

Setup mode is used to control the audible and visible animal alarm
limits, the breath beep pitch, the calendar, and the internal clock.
In setup mode, the Alarm Limit Review button is used to make the menu
selections, and the Advance button is used to change the value.
18
display is lit.
2
Using the 9847V Pulse Oximeter
sensor.
2
level.
2
sensor.
2
sensor fault.
2
Keypad
detector will
2
On / Standby
Advance /
Breath Beep
Volume Control
Alarm Limit
Review
Audible Alarm
Disabled
Indicator (Yellow)
Audible Alarm
Disable
Low Battery
(LED) Indicator
(Yellow)

Advertisement

Table of Contents
loading

Table of Contents