13. Settings
Inverting
The activation mode and the actual contact status resulting from the relay
type and addressing mode are displayed in the next lines.
Status
Contact
Via the keypad or the interface relays can be manually activated to relay vari-
ant ´Driven´. (see Manual 6.10.10)
Relay
Trigger inputs
For the purpose of controlling the measuring sequence two trigger inputs are
provided at ports 8 and 9 (keypad or optocoupler).
The trigger sources ´Key´ and / or ´Optocoupler´ can be defined directly in the
RTA3 by selecting the trigger port
/
... and
PROG
abled altogether by means of ´Off´.
The following trigger functions can be programmed as variants :
Start / stop
Manual
Delete Max / Min
Print the function
Start / stop, level-triggered
Set measured value to zero
If these standard functions prove insufficient, it is also possible to enable
´Macro execution´; any one of the 5 available macros can then be selected
and, in the last line of the macro, interface commands can be entered as re-
quired, separated by ´|´. (see Manual 6.6.5).
Macro 5
Macro 6
Macro 7
Macro 8
Macro 9

13.5.3 Analog output

At sockets A1 and / or A2 it is possible to use not only relays and trigger inputs but
also analog outputs, either as optional ports (4 to 7) with V6 relay trigger adapter
ZA 8006-RTA3 or separately as V5 recording cable ZA 1601-RK. (see Manual
5.1.1) They provide the following output signals :
Recording cable ZA1601-RK
Relay trigger adapter ZA8006-RTA3
or optionally
Various configurations are also possible with analog outputs in the menu ´Output
60
ALMEMO
active / inactive
open / closed
or the trigger function can, for safety reasons, be dis-
Start / stop a measuring operation, edge-controlled
Once-only manual measuring channel scan
Delete all maximum / minimum values
Print the measured value
Start / stop a measuring operation, level-controlled
Set measured value to zero
Execute macro 5
Execute macro 6
Execute macro 7
Execute macro 8
Execute macro 9
Voltage -1.2 to +2.0 V
Voltage 0 to 10 V
Current 0 to 20 mA
®
710
activate
and then using keys
,
,
PROG
PROG
0.1 mV / digit
0.5 mV / digit
1 µA / digit