10. Data logger
dated since the last scan; i.e. in a short scan cycle only high-speed sensors will
appear for most of the time, the slower ones being added at larger intervals.
This method ensures that the measuring instrument can adapt to a wide variety
of sensors - without duplicating a lot of unwanted data.
The device can be further adapted to the specific needs of the user's own ap-
plications by selecting the most suitable output cycle.
The device will automatically offer its own suggestions in a selection box; (D7
options are only available if a D7 sensor is connected).
D7 minimum time By selecting the minimum time a high-speed D7 sensor can
D7 optimum time Optimum time is the shortest cycle that can, with the current
Conversion time
Scan time
D7 maximum time This is the shortest scan cycle in which all measuring chan-
Output per scan cycle
All measured values acquired in a scan cycle can also be output and / or saved
to memory immediately. Both these functions can be activated separately as
and when required via the menu (see man. 6.5.1.3)
Save in scan cycle
Output in scan cycle
28
ALMEMO
be scanned at maximum speed (total sampling rate up to
4000 measuring operations per second).
measured values occur per second, the scan cycle will be
prolonged automatically in increments of 1 ms and some of
the high-speed measuring points may be omitted.
sensor configuration, always be sustained.
This scan cycle corresponds to the conversion time for one
channel; i.e. normally with each cycle one V5 channel and
all updated D7 channels will be scanned and assigned a
time-stamp. However, as soon as all V5 channels have
been scanned at least one special measuring operation is
performed to allow the A/D converter to self-calibrate. If
thermocouples are connected, two additional internal cold
junction measuring operations will also be required.
The scan time is the scan cycle for all standard sensors at
the conversion rate including one special measuring opera-
tion and possibly two cold junction measuring operations
(see above).
nels are always scanned and displayed.
®
710
If even more