Numbering Of Measuring Operations; Memory - Comments Text; Scan Mode - Ahlborn ALMEMO 710 Operating Instructions Manual

Data logger
Hide thumbs Also See for ALMEMO 710:
Table of Contents

Advertisement

measuring operation, a new file will be created; and, if no new file name has
been programmed, the index in the file name extension will automatically be in-
cremented by 1, e.g. ´ALMEMO.002´. Similarly, if the file name now entered al-
ready exists, a new file will be created with the same file name prefix but with a
new index.

10.1.6 Numbering of measuring operations

To identify each measuring operation or each series of measuring operations it
can be assigned a unique number before starting - via ´Settings´ > ´Data log-
ger 2/4´. This number will be output and / or saved with the start of the next
measuring channel scan. When reading out individual measuring operations
these can thus be attributed to certain measuring locations or measuring
points. (see Manual 6.7)
A 6-digit number can be entered via the function ´Number´. (see 9.7)
The number is activated as soon as it has been entered; it will then be followed
in the data logger menu by the letter ´A´ until the next cyclic or manual measur-
ing operation is saved.
To deactivate / delete a number
Number (active)
Number (deleted)

10.2 Memory - comments text

Before starting a measuring operation, an ´Comments text´ can be entered;
this can be up to 64 characters in length and is used to denote its location,
time, and / or subject. This will be saved in the memory configuration and will
appear in the data logger main menu as an alternative to the number (if no
number has been activated).

10.3 Scan mode

Menu ´Data logger´ > menu ´Settings - data logger 2/4´ via function ´Status´ or
via ´Settings´ (page 2) Here in the function ´Scan mode´ you can choose be-
tween 4 different modes with autonomous operation or computer scanning via a
selection box.
Normal Internal cycle or cyclic scanning by the computer
Monitor Internal cycle, not disturbed by computer scans
Failsafe Cyclic scanning by the PC; in the event of PC failure, internal cycle
Sleep
Internal cycle only, automatically switching off for long-term monitoring.
Monitor mode
Monitor mode should be used when a data logger, being operated on a cyclic
basis, is to be monitored occasionally by the computer. Internal cyclic scanning
is not influenced in any way by software scanning; (in WinControl ´Safe initial-
ization´ must be switched off). The internal output cycle is started as soon as
the software starts; it may also have been started previously. When scanning
Measured value acquisition / output
ALMEMO
120001 A
®
710
31

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents