Averaging Over Measuring Channels - Ahlborn ALMEMO 710 Operating Instructions Manual

Data logger
Hide thumbs Also See for ALMEMO 710:
Table of Contents

Advertisement

Set averaging over a cycle
Program the output cycle (see 10.1.3)
Start measuring operation, averaging in progress
Stop the measuring operation
Read out average value / cycle in the function
Averaging over manually set periods of time
Using the same averaging mode but without the cycle the average value can
also be obtained over the period of time from one manual measuring channel
scan to the next.
Set averaging over a cycle
Select the cycle and delete by pressing
Start measuring operation, averaging in progress
Manual measuring channel scan
Average value over time from one measuring channel scan to the next Average value 12.34 ms
 
For recording average values an additional function channel with the
quantity M(t) or the corresponding output function M(t) is required -
instead of the measured value. (see 13.2.5.1, Manual 6.10.4).

14.4.4 Averaging over measuring channels

In all measuring channel scans the average value can also be determined over
a number of associated measuring channels. However, for this average value
a function channel with the measuring range M(n) must be available. (see
13.2.13) If you do not wish to program reference channels and the measuring
channels to be averaged begin with M0.0, you need simply to program the
function channel M(n) to the 2nd channel of the last connector (e.g. M3.1).
(see 13.2.13.1) This will then refer automatically to the series from reference
channel 2 (M0.0) through to reference channel 1 (M3.0 = 1st channel). Other
measuring channel ranges can be activated by programming the reference
channels accordingly. (see 13.2.13.4).
Via ´Select channel´ select
In the next line enter
Freely selectable channel for the
Here the average value M(n) appears
Averaging mode
Output cycle
START
STOP
Average value
Averaging mode
Clr
Cycle timer
START
Manual
average value for meas. channel
up to measuring channel
function channel ´Average value´ M(n)
e.g. average value
ALMEMO
Averaging
CYCL
00:15:00
13.24 m/s
CYCL
00:00:00
10.78 m/s
®
710
71
0.0
4.0
4.1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents