Averaging Over Manually Selected Individual Measuring Opera- Tions; Averaging Over The Full Actual Measuring Duration Or The Fixed Measuring Period - Ahlborn ALMEMO 710 Operating Instructions Manual

Data logger
Hide thumbs Also See for ALMEMO 710:
Table of Contents

Advertisement

When you call up the ´Averaging´ wizard, you will first be faced with a list of
available averaging modes.
13.2.2 Measured value smoothing (see channel functions)
14.4.1 Averaging over manually selected individual measuring operations
14.4.2 Averaging over the full actual measuring duration
or the fixed measuring period
14.4.3 Averaging over a cycle
14.4.4 Averaging over measuring channels
14.4.1 Averaging over manually selected individual measur-
ing operations
To obtain the average over individual measuring operations at particular loca-
tions or times individual measuring channel scans ´E
manually. At all measuring channels whose measured values are to be aver-
aged averaging must be switched to ´CONT´ (continuous) mode.
E1
CLR
MANU
1. To stop a measuring operation if it has been started
2. To set the averaging mode (see 9.7)
To smooth a measured value (if required) select smoothing. Smoothing
3. To delete the average value
The ´Average value´ function displays
The ´Number´ function displays
4. To scan individual measured values ´Ex´ manually
The ´Average value´ function displays
The ´Number´ function displays
14.4.2 Averaging over the full actual measuring duration or
the fixed measuring period
To calculate the average value of all measured values acquired at the scan cy-
cle over a certain period of time the averaging mode for the required measur-
ing channel should be set to ´CONT´.
Averaging can be performed either with or without a cycle.
At the start and stop of any measuring operation a measuring channel scan is
always performed; this ensures that the start value and end value are always
recorded each with the applicable time-of-day. In order to record average value
M
a function channel M(t) is required. (see 13.2.13).
E2
E3
MANU
MANU
´ must be performed
i
E4
M=(
MANU
Stop
Averaging mode
Average value
Number
Manual
Average value
Number
ALMEMO
®
710
Averaging
E
)/N
i
i
CONT
20
0
12.34 m/s
1
69