Operating Instructions ORTHOPHOS XG 3
0123
60 50 681 D3352
D3352.201.05.18.02 04.2016
3

Unit description

3.1

Certification and registration

ORTHOPHOS XG 3 complies with:
with AS/NZS 3200.1.0
● IEC 60601-1
● IEC 60601-1-3
● IEC 60601-2-63
● AS/NZS 3200.1.0
The dental X-ray equipment for extraoral radiography ORTHOPHOS XG
D3352 complies with IEC 60601-2-63
Original language: German
This product bears the CE mark in accordance with the provisions of the
CE mark, general
Council Directive 93/42/EEC of June 14, 1993 concerning medical
devices (MDD).
3.2

Technical data

Unit data
Unit data, power, times
Model designation:
Nominal voltage:
Permissible fluctuation:
Permissible drop under load: 10%
Rated current:
Rated power:
Nominal frequency:
Mains resistance:
Main building fuse:
Power consumption:
Power output of tube
assembly:
Tube voltage:
Tube current:
Maximum setting range:
High-voltage waveform:
High voltage generation
frequency:
Program duration:
Exposure time:
3.1 Certification and registration
ORTHOPHOS XG 3
200 – 240 V
± 10%
12 A
2 kW at 90 kV/12 mA with any radiation
time
50 Hz / 60 Hz
max. 0.8 ohms
25 A slow-blow (16 A for single line)
2 kVA
90 kV/12 mA = 1080 W with any radiation
time
60 – 90 kV (at 90 kV max. 12 mA)
3 – 16 mA (at 16 mA max. 66 kV)
60 kV / 3 mA to 90 kV / 12 mA
High-frequency multipulse
Residual ripple ≤ 4 kV
40 – 120 kHz
See "Program values".