Replacing the
The UV lamp used for curing models has a long, but limited, working life. You 
UV Lamp
can test and adjust its effective power with the UV Calibration wizard (see 
"Testing and Calibrating the UV Lamp" on page 6-63).  If you need to replace 
the UV lamp, follow these instructions: 
1.  Start the UV Lamp Replacement wizard from the Options menu (see 
Warning
Before continuing, make sure that the safety interlock in the printer cover 
is not defeated, and that the lamp is not hot.
2.  Make sure that the build tray is empty, and confirm this in the wizard 
3.  On the UV-assembly cover, loosen (do not remove) the thumbscrew 
4.  Remove the other four (4) thumbscrews that secure the UV-assembly 
DOC-34022 Rev. D
"Starting the Material Replacement wizard" on page 6-14). 
screen. Then, close the printer cover and click Next.
Figure 6-96    "Empty tray" confirmation screen
The axes move to the "home" position.
indicated in the photograph below.
Figure 6-97    UV-assembly cover
cover, and remove the cover.