Final Width = 215mm
10.4-28
Section 10: Calibrating the Model 9100: AC Current Function
10.4.6

AC Current Calibration

10.4.6.1
Introduction
This section is a guide to calibrating the Model 9100's AC Current Function using its front
panel controls. The following topics are covered:
10.4.6.2 Calibration Equipment Requirements
10.4.6.3 Interconnections
10.4.6.4 Calibration Setup
10.4.6.5 Calibration Procedure
10.4.6.2
Calibration Equipment Requirements
A traceably characterized, long scale-length, Standards DMM connected to the 9100
output terminals by a short, high-quality coaxial cable. For example, a Fluke Model
1281 Digital Multimeter. Note that some DMMs may require a precision AC Current
shunt to measure up to 10A.