Fluke 9100 User Handbook Manual page 124

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

Final Width = 215mm
7.19.4 Auxiliary AC Voltage (Representing AC Current Output) Accuracy
Output Current
0.32A to 3.20A
3.21A to 32.00A
32.1 to 320.0A @ 1mV/Amp
321A to 3.2KA @ 1mV/Amp
0 to 160.0A @ 0.2mV to 1mV/Amp
160A to 3.2kA @ 0.2mV to 1mV/Amp
>3.2kA
NOTES: Improved specs are available at intermediate points depending on scaling factor.
Auxiliary AC voltage is available as a simulation of the output of a resistive shunt, and can therefore be expressed as being equivalent to Amps.
Also directly available as a voltage harmonic output. When used as a harmonic with n>1, add 0.04% to percentage of output specifications.
7.19.5 Other Voltage Waveshapes
All subject to 65Hz/1kHz maximum as defined in main Model 9100 specification tables.
Note that all waveshapes are specified to 4.525 pk only (4 ranges; equivalent to the 3.2V sinewave range peak).
Functionality above this limit is available but unspecified.
The 4.525 mV pk range is allowed to function down to zero.
Specifications as main 9100 ranges.
7.19.6 DC Current Accuracy
Output Current
Accuracy
(% Output + Floor)
0 to 320.0mA
0.016 + 9.6 A
0.320A to 3.200A
0.06 + 118 A
3.20A to 10.50A
0.055 + 940 A
10.50A to 20.00A
0.055 + 4.5mA
3.200 to 32.00A, 10t coil
0.26 + 1.18mA
32.00 to 200.0A, 10t coil
0.255 + 45mA
16.00 to 160.0A, 50t coil
0.26 + 5.9mA
160.0 to 1000A, 50t coil
0.255 + 225mA
7-28
Section 7: Model 9100: Specifications
Frequency
Accuracy
Compliance Current
Band
(% Output + Floor)
mA (O/P Impedance)
10 - 3K
0.12% + 0.01A
10
10 - 3K
0.12% + 0.01A
10
10 - 3K
0.04% + 0.02A
<20mA
10 - 3K
0.04% + 0.2A
<20mA
10 - 3K
0.04% + 0.1A
<20mA
10 - 3K
0.04% + 1A
<20mA
10 - 3K
0.06% + 2A
<7mA @ 7V