Dc Voltage Calibration - Fluke 9100 User Handbook Manual

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

Final Width = 215mm
10.4-6
Section 10: Calibrating the Model 9100: DC Voltage Function
10.4.3

DC Voltage Calibration

10.4.3.1
Introduction
This section is a guide to calibrating the Model 9100's DC Voltage Function using its front
panel controls. The following topics are covered:
10.4.3.2 Calibration Equipment Requirements
10.4.3.3 Interconnections
10.4.3.4 Calibration Setup
10.4.3.5 Calibration Procedure
10.4.3.2
Calibration Equipment Requirements
A traceably characterized, long scale-length, Standards DMM connected to the Model
9100's output terminals by short, high-quality leads. For example, a Fluke Model 1281
Digital Multimeter.