The 9100 first checks for the presence of the correct SRAM card. If there is no card in
the slot, if the card in the slot is not a SRAM card, or if it is a SRAM card but not write-
enabled, then the following screen is displayed:
Card slot test.
Insert a write enabled STATIC RAM
card into the selected slot.
(If no card in selected slot, or if
card not write-enabled, or if not a
TODAYS DATE
E X I T
After correcting the defect, press the OK screen key. This transfers to a new screen, and
the sequence of tests begins. The test in progress is reported on the screen:
The 9100 first checks the state of the card's internal battery. If the battery voltage is low,
this will be stated on the screen:
Card slot test.
Battery level is low.
TODAYS DATE
If there is no battery in the card, or if the battery cannot support read/write operations,
then the statement will be '
Battery level is dead.
Note:
For a re-chargeable card with a low battery, the low/dead message may take several
minutes to clear after pressing
OK
.
Section 8: Model 9100 — Routine Maintenance and Test
SRAM card)
TIME
O K
TIME
'.
Continued Overleaf