7.8

Conductance Specifications

7.8.1
Conductance Accuracy
Conductance Output
1Year — Tcal 5 C
UUTi Low & High
2.5nS - 25.0nS
0.40
25.0nS - 250.0nS
0.20
250.0nS - 2.5 S
0.12
2.5 S - 25.0 S
0.05
25.0 S - 250.0 S
0.05
250.0 S - 2.5mS
0.04
7.8.3
Source Current — Limits
Hardware Configuration Limits
on Span of Output Conductance
02.5000 nS to 25.0000 nS
025.001 nS to 250.000 nS
0.25001 S to 2.50000 S
02.5001 S to 25.0000 S
025.001 S to 250.000 S
0.25001 mS to 2.50000 mS
7.8.4
Other Conductance Specifications
Maximum Measurement Voltage:
Settling Time:
4-wire Lead Compensation:
NOTES: [1]
Tcal = temperature at calibration. Factory calibration temperature = 23 C
7.8.2
Conductance Span
Accuracy
(% Output)
[1]
UUTi Super
N/A
0.45
0.27
0.12
0.12
0.09
Current Source Limits
UUTi Low
UUTi High
4nA to 32nA
2.5nA to 200nA
8nA to 32nA
25nA to 350nA
25nA to 320nA
250nA to 3.5 A
250nA to 3.2 A
2.5 A to 35 A
2.5 A to 32 A
25 A to 350 A
25 A to 320 A
250 A to 3.5mA
10V;
I
/
:
10V
measure
actual
to within 10% of accuracy :
2.5nS - 250nS :
250nS - 25 S
:
25 S - 2.5mS :
Max total lead resistance :
50
Nominal lead resistance rejection: 10000:1
Section 7: Model 9100: Specifications
Resolution
vs
Absolute
Span of Values
Resolution
0.1pS
02.5000 nS
to 25.0000 nS
1pS
002.500 nS
to 250.000 nS
10pS
0.00250 S
to 2.50000 S
100pS
00.0025 S
to 25.0000 S
1nS
000.002 S
to 250.000 S
10nS
0.00001 mS to 2.50000 mS
UUTi Super
N/A
250nA to 3.5 A
2.5 A to 35 A
25 A to 350 A
250 A to 3.5mA
2.5A to 35mA
<1s
<0.3s
<0.08s