Fluke 9100 User Handbook Manual page 257

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

10.4.8.7
Calibration of 'UUTi Super' Source Current Hardware Configurations
For calibration of UUTi Super source current hardware configurations use Table 10.4.8.2.
Table 10.4.8.2: Capacitance UUTi Super Hardware Configurations and Calibration Points
Output span covered
Suitable output to
by hardware configuration
select hardware
configuration
40.001 F to 400.000 F
0.40001mF to 4.00000mF
4.0001mF to 40.0000mF
Calibration Targets
Recommended
100 F
60.000 F
190.000 F
1mF
0.60000mF
1.90000mF
10mF
6.0000mF
19.0000mF
Section 10: Calibrating the Model 9100: Capacitance Function
Minimum
Maximum
41.000 F
80.000 F
250.000 F
390.000 F
0.41000mF
0.80000mF
2.50000mF
3.90000mF
4.1000mF
8.0000mF
25.0000mF
39.0000mF
Factor
Number
1
2
1
2
1
2
10.4-49