Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

Final Width = 215mm
6-6
Section 6: 9100 System Operation — SCPI Language
6.3.2

Interconnections

Instruments fitted with an IEEE 488
interface communicate with each other
through a standard set of interconnecting
cables, as specified in the IEEE 488.1
Standard document.
The IEEE-488 interface socket, J101, is
fitted on the rear panel. It accommodates
the specified connector, whose pin
designations are also standardized as
shown in Table 6.2
Pin
No.
Name
Description
1
DIO 1
Data Input/Output Line 1
2
DIO 2
Data Input/Output Line 2
3
DIO 3
Data Input/Output Line 3
4
DIO 4
Data Input/Output Line 4
5
EOI
End or Identify
6
DAV
Data Valid
7
NRFD
Not Ready For Data
8
NDAC
Not Data Accepted
9
IFC
Interface Clear
10
SRQ
Service Request
11
ATN
Attention
12
SHIELD
Screening on cable (connected to 9100 safety ground)
13
DIO 5
Data Input/Output Line 5
14
DIO 6
Data Input/Output Line 6
15
DIO 7
Data Input/Output Line 7
16
DIO 8
Data Input/Output Line 8
17
REN
Remote Enable
18
GND 6
Gnd wire of DAV twisted pair
19
GND 7
Gnd wire of NRFD twisted pair
20
GND 8
Gnd wire of NDAC twisted pair
21
GND 9
Gnd wire of IFC twisted pair
22
GND 10
Gnd wire of SRQ twisted pair
23
GND 11
Gnd wire of ATN twisted pair
24
GND
9100 Logic Ground (internally connected to Safety Ground)
Table 6.2 Connector J101 - Pin Designations