10.4.6.4
Calibration Setup
1. Connections
Ensure that the Model 9100 is connected to the Standards DMM as shown in Fig. 10.4.6.1 or 10.4.6.2
(depending on the current being calibrated), and that both instruments are powered on and warmed up. If the
measuring instrument has a switchable signal guarding facility it should be switched to remote guard. If it
does not have a guard facility, there can be no guard lead to the DMM.
2. 9100
a. Ensure that the 9100 is in STANDARD CAL mode and then select the AC Current function by pressing
the 'A' key on the right of the front panel, followed by pressing the ' A' screen key adjacent to the display.
b. Ensure that Sinusoidal waveshape is selected, and that the Phase facility is deselected (
lettering on light background).
Table 10.4.6.1: AC Current (Zero to 320mA) Hardware Configurations and Calibration Points
Output span
Output
covered by
to select
hardware
hardware
configuration
config.
Default
000.000 A
100 A
300.000 A
to
300.000 A
320.000 A
300.000 A
0.32001mA
1mA
3.00000mA
to
3.00000mA
3.20000mA
3.00000mA
3.2001mA
10mA
30.0000mA
to
30.0000mA
32.0000mA
30.0000mA
32.001mA
100mA
300.000mA
to
300.000mA
320.000mA
300.000mA
Table 10.4.6.2: AC Current (0.32A to 20A) Hardware Configurations and Calibration Points
Output span
Output
covered by
to select
hardware
hardware
configuration
config.
Default
0.32001A
1A
3.00000A
to
3.00000A
3.20000A
3.00000A
3.2001A
10A
10.0000A
to
10.0000A
20.0000A
10.0000A
10.4-30
Section 10: Calibrating the Model 9100: AC Current Function
Calibration Targets
Amplitude
Minimum
Maximum
Default
180.000 A
320.000 A
110.00Hz
180.000 A
320.000 A
3.0000kHz
180.000 A
320.000 A
30.000kHz
1.80000mA
3.20000mA
110.00Hz
1.80000mA
3.20000mA
3.0000kHz
1.80000mA
3.20000mA
30.000kHz
18.0000mA
32.0000mA
110.00Hz
18.0000mA
32.0000mA
3.0000kHz
18.0000mA
32.0000mA
30.000kHz
180.000mA
320.000mA
110.00Hz
180.000mA
320.000mA
3.0000kHz
180.000mA
320.000mA
30.000kHz
Calibration Targets
Amplitude
Minimum
Maximum
Default
1.80000A
3.20000A
110.00Hz
1.80000A
3.20000A
3.0000kHz
1.80000A
3.20000A
10.000kHz
7.0000A
10.5000A
110.00Hz
7.0000A
10.5000A
3.0000kHz
7.0000A
10.5000A
10.000kHz
label dark
Factor
Number
Frequency
Minimum
Maximum
10.00Hz
320.00Hz
1
1.0000kHz
5.000kHz
2
20.000kHz
30.000kHz
3
10.00Hz
320.00Hz
1
1.0000kHz
5.000kHz
2
20.000kHz
30.000kHz
3
10.00Hz
320.00Hz
1
1.0000kHz
5.000kHz
2
20.000kHz
30.000kHz
3
10.00Hz
320.00Hz
1
1.0000kHz
5.000kHz
2
20.000kHz
30.000kHz
3
Factor
Number
Frequency
Minimum
Maximum
10.00Hz
320.00Hz
1
2.0000kHz
4.000kHz
2
7.000kHz
10.000kHz
3
10.00Hz
320.00Hz
1
2.0000kHz
4.000kHz
2
7.000kHz
10.000kHz
3