Fluke 9132 User Manual

Portable ir calibrator
Hide thumbs Also See for 9132:

Advertisement

Quick Links

9132
Portable IR Calibrator
User's Guide
Rev. 672501 ENG
Fluke-Direct
.com
info@Fluke-Direct.com
1.888.610.7664

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 9132

  • Page 1 9132 Portable IR Calibrator User’s Guide Rev. 672501 ENG Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 2 Limited Warranty & Limitation of Liability Each product from Fluke Corporation, Hart Scientific Division (“Hart”) is warranted to be free from de- fects in material and workmanship under normal use and service. The warranty period is one year for the Portable IR Calibrator.
  • Page 3: Table Of Contents

    Front Panel ...................12 General Operation ..............15 5.1 Changing Display Units ...............15 5.2 Switching to 230 V Operation ..............15 Controller Operation .............. 17 6.1 Well Temperature .................17 6.2 Temperature Set-point .................17 6.2.1 Programmable Set-points ................17 6.2.2 Set-point Value .................... 19 6.3 Temperature Scale Units ..............19 6.4 Scan .....................19 6.4.1 Scan Control ....................19 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 4 Calibration Procedure ............29 8.1 Calibration Points .................29 8.2 Calibration Procedure ................29 8.2.1 Compute DELTA ..................29 8.2.2 Compute R0 and ALPHA ................30 8.2.3 Accuracy and Repeatability ................ 31 Maintenance ................33 10 Troubleshooting ..............35 10.1 Troubleshooting Problems, Possible Causes, and Solutions ....35 10.2 Comments ....................36 10.2.1 EMC Directive ..................... 36 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 5 10.2.2 Low Voltage Directive (Safety) ..............36 Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 6: Before You Start

    Before You Start Introduction The Hart Scientific 9132 Portable IR Calibrator may be used as a portable instrument or bench top temperature calibrator for calibrating point IR thermometers. The 9132 is small enough to use in the field, and accurate enough to use in the lab. Calibrations may be done over a range of 50°C to 500°C (122°F to 932°F).
  • Page 7 9132 Portable IR Calibrator User’s Guide Table 1 Symbols Symbol Description AC (Alternating Current) AC-DC Battery Complies with European Union directives Double Insulated Electric Shock Fuse PE Ground Hot Surface (Burn Hazard) Read the User’s Guide (Important Information) Fluke-Direct .com info@Fluke-Direct.com...
  • Page 8: Safety Information

    Always replace the fuse with one of the same rating, voltage, and type. ● This instrument is intended for indoor use only. ● Overhead clearance is required. DO NOT place this instrument under a cabinet or other ● structure. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 9: Cautions

    9132 Portable IR Calibrator User’s Guide DO NOT use this unit for any application other than calibration work. ● DO NOT use this unit in environments other than those listed in the user’s guide. ● DO NOT operate near flammable materials.
  • Page 10: Specifications And Environmental Conditions

    4” H x 6” W x 7” D (102 x 152 x 178 mm) Weight 4 lb. (1.8 kg) Temperature differences between the IR surface and the reference probe are calculated as shown in the graph in Figure 1. 9132 Gradient between probe and IR surface -0.1 -0.2 -0.3 -0.4...
  • Page 11: Environmental Conditions

    9132 Portable IR Calibrator User’s Guide NOTE: As the calibrator temperature increases, the difference between the temperature sensed by the reference thermometer and the temperature of the face of the target increases. Environmental Conditions Although the instrument has been designed for optimum durability and trouble-free operation, it must be handled with care.
  • Page 12: Quick Start

    Unpacking Unpack the calibrator carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier immediately. Verify that the following components are present: 9132 Calibrator ● Power Cord ● User’s Guide ●...
  • Page 13 9132 Portable IR Calibrator User’s Guide When the set-point temperature is changed the controller switches the well heater on or off to raise or lower the temperature. The displayed well temperature gradually changes until it reaches the set-point temperature. The well may require 5 to 10 minutes to reach the set-point depending on the span.
  • Page 14: Parts And Controls

    115 V. This action will cause the fuses to blow and may damage the instrument. Serial Port – A DB-9 male connector is present for interfacing the calibrator to a computer or terminal with serial RS-232 communications. Fluke-Direct .com info@Fluke-Direct.com...
  • Page 15: Front Panel

    9132 Portable IR Calibrator User’s Guide Fan – The fan inside the calibrator has two speeds and runs continuously when the unit is being operated to provide cooling for the instrument. The fan runs slow for heating and maintaining operation and runs fast for rapid cooling. Slots are provided for airflow. The area around the calibrator must be kept clear to allow adequate ventilation.
  • Page 16 EXIT – Used to exit a function and to skip to the next function. Any changes made to the dis- played value are ignored. Target Assembly – The target assembly is 57 mm (2.25 in) in diameter and has an emisivity of 0.95. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 17: General Operation

    Press the “UP” or “DOWN” key to change units. Switching to 230 V Operation The 9132 is switchable from 115 V ac to 230 V ac 50/60 Hz. Switching the voltage can change the calibration. Therefore, we recommend that the instrument be calibrated after the voltage is switched.
  • Page 18: Controller Operation

    Well temperature in degrees Celsius 100.0 C Access set-point memory Set-point memory 1, 100°C currently used. 1. 100. To change to another set-point memory press “UP” or “DOWN”. New set-point memory 4, 300°C 4. 300. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 19 9132 Portable IR Calibrator User’s Guide Figure 4 Controller Operation Flowchart Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 20: Set-Point Value

    The scan is controlled with the scan on/off function that appears in the main menu after the temperature scale units. Scan function off Sc=OFF Press “UP” or “DOWN” to toggle the scan on or off. Scan function on Sc=On Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 21: Scan Rate

    9132 Portable IR Calibrator User’s Guide Press “SET” to accept the present setting and continue. Accept scan setting 6.4.2 Scan Rate The next function in the main menu is the scan rate. The scan rate can be set from .1 to 99.9°C/ min.
  • Page 22: Proportional Band

    Access heater power in secondary menu Flashes SEC for secondary menu and then displays the heater power Heater power in percent 12.0P Access proportional band Flashes PROP for secondary menu and then displays the proportional band PROP setting Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 23: Controller Configuration

    9132 Portable IR Calibrator User’s Guide Proportional band setting To change the proportional band press “UP” or “DOWN”. New proportional band setting 10.0 To accept the new setting press “SET”. Press “EXIT” to continue without storing the new value. Accept the new proportional band setting 6.10...
  • Page 24: 6.12.1 Baud Rate

    The final parameter in the serial interface menu is the linefeed mode. This parameter enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return. The linefeed parameter is indicated by, Fluke-Direct .com info@Fluke-Direct.com...
  • Page 25: Calibration Parameters

    6.13.3 DELTA This probe parameter characterizes the curvature of the resistance-temperature relationship of the sensor. The value of this parameter is set at the factory for best instrument accuracy. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 26: Digital Communications

    SERIAL. Finally press “SET” to enter the serial parameter menu. In the serial interface parameters menu are the baud rate, the sample rate, the duplex mode, and the linefeed parameter. Figure 5 RS-232 Cable Wiring Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 27: Serial Operation

    9132 Portable IR Calibrator User’s Guide 7.2.1.1 Baud Rate The baud rate is the first parameter in the menu. The display prompts with the baud rate pa- rameter by showing “bAUd”. Press “SET” to choose to set the baud rate. The current baud rate value is displayed.
  • Page 28 Spaces may be added within command strings and will simply be ignored. Backspace (BS, SCII 8) may be used to erase the previous character. A terminating CR is implied with all commands. Fluke-Direct .com info@Fluke-Direct.com...
  • Page 29 9132 Portable IR Calibrator User’s Guide Table 3 Serial Commands Command Command Returned Acceptable Command Description Format Example Returned Example Values Display Temperature Read current set-point s[etpoint] set: 999.9 {C or F} set: 100.00 C Set current set-point to n s[etpoint]=n s=200.0...
  • Page 30: Calibration Procedure

    Repeat step 2 for the other two set points recording them as T and R and T and R respectively. Using the recorded data, calculate new values for the R0, ALPHA, and DELTA parameters using the following equations. 8.2.1 Compute DELTA Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 31: Compute R0 And Alpha

    9132 Portable IR Calibrator User’s Guide - Measured temperature using a thermometer – Value of the set-point resistance from the display. (Press SET and DOWN at the same time.) Where: and R are the measured temperature and resistance at 50.0°C and R are the measured temperature and resistance at 200.0°C...
  • Page 32: Accuracy And Repeatability

    8 Calibration Procedure 8.2.3 Accuracy and Repeatability Check the accuracy of the instrument at various points over the calibration range. If instrument does not pass specification at all set-points, repeat the Calibration Procedure. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 33: Maintenance

    If the instrument is used in a manner not in accordance with the equipment design, the ● operation of the instrument may be impaired or safety hazards may arise. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...
  • Page 34: Troubleshooting

    The display shows -init-, then displays 9132, and then displays the firmware version. After performing the master reset sequence, all of the configuration parameters are reset to their default values. Reprogram the R0, ALPHA, and DELTA parameters into the 9132 memory (see Section 6.13, Calibration Parameters) and any other applicable configuration parameters.
  • Page 35: Comments

    10.2.2 Low Voltage Directive (Safety) In order to comply with the European Low Voltage Directive (73/23/EEC), Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN 61010-1) and the IEC 1010-2-010 (EN 61010-2-010) standards. Fluke-Direct .com info@Fluke-Direct.com 1.888.610.7664...