10.2.3.1
Calibration Enable Switch
The Calibration Enable DIP-switch is accessible, using a small screwdriver, through a
recess on the Model 9100's rear panel.
Cal Key Disabled
With the Calibration Enable switch in the 'DISABLE' position, any
attempt to select Calibration mode by pressing the 'CALIB' screen
softkey will result in the screen prompt "Calibration switch not
enabled!" being displayed, and access to the calibration mode will
be denied.
Cal Key Enabled
With the Calibration Enable switch in the 'ENABLE' position,
pressing the 'CALIB' screen key will result in the 'Password Entry
for Calibration' user prompt screen being displayed.
Password Entry
For Calibration
Enter your password:
TODAYS DATE
E X I T
10.2.3.2
Password
Before the Calibration mode menu screen can be displayed you must now enter a valid
password using the Model 9100's alphanumeric keyboard. For information about the
initial 'shipment' password, and about the method of changing this to a custom password,
refer to Section 3.3.2.23 of the Model 9100 Universal Calibration System User's
Handbook - Volume 1 - Operation.
As each character in the password is entered, security code icons will appear on the screen
as shown above. Once the correct password has been entered, pressing the '
key will result in the Calibration Mode menu screen being displayed. If the password is
incorrect, the error message "PASSWORD INCORRECT" will be displayed and the
security code character icons will disappear, enabling you to attempt correct password
entry again.
The '
' screen softkey will take you back to the Mode Selection screen.
EXIT
Section 10: Calibrating the Model 9100: Calibration Mode