Final Width = 215mm
6.6.5

STATus Subsystem

This subsystem is used to enable bits in the Operation and Questionable Event registers. The Operation and Questionable: Event,
Enable and Condition registers can be interrogated to determine their state. For further information regarding the Status structure,
refer to Sub-Section 6.5.4.
6.6.5.1
STATus Subsystem Table
Keyword
STATus
:OPERation
[:EVENt]?
:ENABle(?)
:CONDition?
:QUEStionable
[:EVENt]?
:ENABle(?)
:CONDition?
:PRESet
6.6.5.2
STAT:OPER[:EVEN]?
Purpose
OPER? returns the contents of the Operation Event register, clearing the register.
Response
A <DNPD> in the form of an Nr1 number is returned. The value of the number, when converted to base 2 (binary), identifies the
Operation Event register bits to determine their current status.
For example (refer to Fig. 6.2):
If the 9100 had just performed a selftest, the 'TESTING' bit 8 of the register would be set, and if no other Operation Event bits
were enabled, the number 256 would be returned. Bit 8 (indeed, all bits in the register) would be reset by this query.
6-50
Section 6: 9100 System Operation — SCPI Language
Parameter Form
<DNPD>
<DNPD>
Notes
[Query Only]
[Query Only]
[Query Only]
[Query Only]