Fluke 9100 User Handbook Manual page 229

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

10.4.5.3
Interconnections
Interconnections for Low Current Outputs ( 1A)
DANGER
HIGH VOLTAGE
sHi
sLo
!
Hi
Lo
1500Vpk
max
I+
I-
Pin 8
Pin 7
Keep Coaxials
Close
Together
Guard
I+
!
Hi
Fig. 10.4.5.1 DC Low Current Calibration —
Interconnections
Interconnections for High Current Outputs (>1A)
15Vpk
max
!
15Vpk
max
I+
Shunt
e.g. 4953
Hi
Standards DMM
Guard
I-
Lo
Reset
Fig. 10.4.5.2 DC High Current Calibration —
Section 10: Calibrating the Model 9100: DC Current Function
HIGH VOLTAGE
DANGER
sHi
sLo
15Vpk
max
!
!
Hi
Lo
1500Vpk
max
I+
I-
15Vpk
max
I-
Case
Lo
Standards DMM
Guard
Guard
I+
I-
!
Hi
Lo
Reset
Interconnections
Final Width = 215mm
10.4-21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents