Fluke 9100 User Handbook Manual page 149

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

8.3.4.2
Display Memory Checks
Pressing the DISPLAY MEMORY key on the 'Select test' menu screen transfers to the
'Memory test.' screen, and the sequence of tests begins. The test in progress is reported
on the screen:
D I S P L A Y
Selftest.
M E M O R Y
K E Y B R D
Select test using softkeys
D I S P L A Y
M E M O R Y
C A R D
T R A C K E R
P R I N T E R
TODAYS DATE
TIME
E X I T
The other tests reported are:
Performing WALKING ONES test 1.
Performing READ/WRITE test 2.
Performing WALKING ONES test 2.
The result of the whole Memory test is reported once testing is complete:
Memory test.
Display memory PASSED.
TODAYS DATE
TIME
E X I T
If a failure is reported, rectification will require access to the internal circuitry, so no
further user action is recommended, except to report the result to your Fluke Service
Center.
EXIT
returns to the Interface '
Select test
Section 8: Model 9100 — Routine Maintenance and Test
Memory test.
Performing READ/WRITE test 1.
TODAYS DATE
TIME
Memory test.
Display memory FAILED.
TODAYS DATE
TIME
E X I T
' menu screen.
Final Width = 215mm
8-17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents