Table of Contents

Advertisement

4. Re-Program the Firmware:
a. Switch 9100 Power ON.
b. Observe the following growth pattern on the 9100 LCD display:
The process takes approximately 1 minute 10 seconds to complete. When complete, a
repetitive, pulsed audible tone will be heard.
If for any reason a continuous audible tone is heard, the update has not been successful.
Note the current state of the growth pattern on the display, and relay this information
back to the Service Center.
5. Recover the Operational State of the 9100:
a. Switch 9100 Power
OFF
.
b. Return
Switch 6
to the
DISABLE
switches.
c. Replace and secure the switch cover.
d. Remove the PCMCIA card, switch the 9100 Power
1 minute until the Power-On Selftest is complete.
e. On the right of the 9100 front panel, press the
f. At the bottom of the 9100 LCD display, press the soft
that the firmware issue, shown on the screen, matches that on the PCMCIA card.
g. Transfer back to the Mode Selection screen by pressing the
6. Full Self Test (Refer to paras 8.3.1 & 8.3.2):
a. Execute a Full Self Test and record any errors. If printing facilities are available,
select the PRINT option to obtain a hard copy of the results (refer also to paras
8.4).
b. Note any differences between the self tests at items 1 and 6.
c. Report the results of the upgrade procedure back to the Service Center.
The Model 9100 firmware upgrade is now complete. Please return your PCMCIA to the
Service Center.
Section 8: Model 9100 — Routine Maintenance and Test
position. Do not disturb any of the other
and wait for approximately
ON
Mode
key.
key, and check
CONFIG
key.
Mode
Final Width = 215mm
8-5