Fluke 9100 User Handbook Manual page 234

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

Final Width = 215mm
Table 10.4.5.3: Auxiliary DC Current Calibration Points
Hardware Spans
0.000mA to 320.0mA
-300.000mA
(one range)
+300.000mA
0.321A to 3.200A
+3.00000A
3.21 to 32.00A
+10.0000A
10.4-26
Section 10: Calibrating the Model 9100: DC Current Function
Calibration Targets
Default
Minimum
-320.000mA
+180.000mA
-3.00000A
-3.20000A
+1.80000A
-10.0000A
-10.5000A
+7.00000A
Maximum
-180.00mA
+320.000mA
-1.80000A
+3.20000A
-7.00000A
+10.5000A