8.A.2.2.9
Device-Dependent Errors Reported both Locally and Remotely
Calibration (continued)
4060,"Corrupt offset DAC factor"
4061,"Using default DAC factors"
-7021,"Temperature must be in the range
00.00 - 50.00"
-7022,"A valid Cold Junction temperature
is required"
Characterization
4501,"Limits: main DAC gain"
4502,"Limits: composite DAC zero"
4503,"Limits: trim DAC gain"
4504,"Limits: offset DAC gain"
4505,"Limits: main DAC linearity"
4506,"Failed to write to flash RAM"
4507,"Limits: gain of 2 zero"
4508,"Limits: gain of 0.5 zero"
4509,"Limits: DAC output zero"
4510,"Limits: 0.75 buffer zero"
4511,"Limits: DAC positive zero"
4512,"Limits: DAC negative zero"
4513,"Limits: DAC positive FR"
4514,"Limits: DAC negative FR"
4515,"Limits: DAC \361FR ratio"
4516,"Limits: DAC max - DAC min"
4517,"Limits: resistor ratios"
4518,"Failed to save resistor ratios"
4519,"Limits: main DAC offset"
4520,"Failed to save impedance offset"
4521,"LF AC Chrctn impossible: default set"
4522,"Excess LF AC flatness"
4523,"Failed to save DDS lfac error"
4524,"Limits: Gain of 1 zero"
4525,"Limits: PWR DAC gain"
4526,"Limits: PWR trim DAC gain"
4527,"Limits: PWR DAC zero"
4528,"Limits: PWR DAC linearity"
ALWAYS: record the total message content for possible use by the Service Center.
8-A6
Appendix A to Section 8: 9100 Maintenance — Error Reporting Subsystem
(Contd.)
DAC Compensation
5001,"Corrupt main DAC gain"
5002,"Corrupt trim DAC gain"
5003,"Corrupt composite DAC zero"
5004,"Corrupt lookup table"
5005,"Corrupt Vmax. Vmin"
5006,"Corrupt +ve zero (DAC)"
5007,"Corrupt -ve zero (DAC)"
5008,"Corrupt polarity gain (DAC)"
5009,"Corrupt LFAC correction"
5011,"Gain request limited"
5012,"Failed to read from flash RAM"
Cold Junction Compensation
5020,"Invalid CJC measurement -
default to 25 C"
5021,"Temperature table: no entry found"
Configuration
4002,"Failed to save configuration"
-7003,"Day entry is not a valid number"
-7004,"Day separator is incorrect"
-7005,"Month entry is not a valid number"
-7006,"Month separator is incorrect"
-7007,"Century entry is not a valid
number"
-7008,"Year entry is not a valid number"
-7009,"Year separator is incorrect"
-7010,"Month entry is not a valid month"
-7011,"Day entry is not a valid day"
-7012,"Hours entry is not a valid number"
-7013,"Minutes entry is not a valid
number"
-7014,"Entry does not give a valid time
setting"
-7017,"Safety Voltage must be in the
range 10.00 V -110.000 V"
-7031,"The transducer scaling factor must
be between 45 uV and 10 mV