6.7.4
Harmonic Selection
This sub-system defines the major hardware configuration of the power option harmonic feature.
6.7.4.1
[SOURce]:HARMonic(?)<dnpd>
This command selects the harmonic frequency on the auxiliary channel.
• The range of the <dnpd> is 1.0 to 40.0. Values outside this will generate a Data out of range error.
6.7.4.2
[SOURce]:HARMonic:MAIN:VOLTage[:LEVel][:IMMediate][:AMPLitude](?)
This sets the voltage amplitude, in RMS of the main channel.
6.7.4.3
[SOURce]:HARMonic:AUXiliary:CURRent[:LEVel][:IMMediate][:AMPLitude](?)
<dnpd>
This sets the current amplitude, in RMS, of the auxiliary channel (Note the main channel is voltage only).
6.7.4.4
[SOURce]:HARMonic:AUXiliary:VOLTage[:LEVel][:IMMediate][:AMPLitude](?)
<dnpd>
This sets the voltage amplitude, in RMS of the auxiliary channel.
6.7.4.5
[SOURce]:HARMonic:PANGle(?)<dnpd>
This command sets the angular difference between the main and auxiliary waveforms. It is set as a phase angle in degrees.
• If the instrument is not in harmonics, then a settings conflict error will be generated.
• The range of the PANGle<dnpd> is +180.0 to -180.0. Values outside this will generate a Data out of range error.
Section 6: 9100 System Operation — SCPI Language
<dnpd>
6-61