Dc Voltage Ranges Calibration Adjustment - Fluke 6003A Operator's Manual

Three phase power calibrator
Hide thumbs Also See for 6003A:
Table of Contents

Advertisement

DC Voltage Ranges Calibration Adjustment

To adjust the calibration of the dc voltage ranges:
1. Select Voltage DC#1 from the Calibration Menu.
2. Push the Select softkey to confirm.
3. Switch the Product output terminals to Operate.
4. Follow the instructions provided on the Product display and Table 5-4 to
adjust the Product output for the calibration adjustment points.
5. To adjust the Product output to the calibration adjustment points, push the
Select softkey and use ,,, and , the rotary knob, or numeric keyboard
to adjust the output to the nominal value.
6. Push the Write softkey to confirm the new value. To skip a calibration
adjustment point where calibration has already been entered, push the Exit
softkey.
7. Repeat steps 1 through 6 for Voltage DC#2 and Voltage DC#3.
Nominal Value [V]
1 V
10 V
-1 V
-10 V
10 V
30 V
-10 V
-30 V
30 V
70 V
-30 V
-70 V
70 V
140 V
-70 V
-140 V
140 V
280 V
-140 V
-280 V
Verification, Calibration Adjustment, and Maintenance
Table 5-4. Voltage DC Function (Voltage DC#x)
Set Limits [V]
100 µV
500 µV
100 µV
500 µV
1 mV
2 mV
1 mV
2 mV
4 mV
4 mV
4 mV
4 mV
7 mV
7 mV
7 mV
7 mV
15 mV
15 mV
15 mV
15 mV
Complete Calibration Adjustment Procedure
Range [V]
10 V
10 V
-10 V
-10 V
30 V
30 V
-30 V
-30 V
70 V
70 V
-70 V
-70 V
140 V
140 V
-140 V
-140 V
280 V
280 V
-280 V
-280 V
5
5-25