Measurement Time and Resolution

The table below lists the measurement time configuration options available
for the multimeter, for each measurement function.

DC, Resistance and Temperature Measurements

For dc voltage, dc current, resistance and temperature measurements
only, the multimeter allows for two ways to set integration time, both of
which affect the measurement resolution. These are in the number of
power–line cycles (NPLCs) and aperture.
Using NPLCs sets the number of power line cycles, during which the
multimeter's analog–to–digital (A/D) converter samples the input signal for
a measurement. The multimeter automatically detects the AC line
frequency, using it and the NPLC selection to set integration time.
Selecting a larger NPLC value yields better resolution; a shorter value
provides for faster measurements. Choices are 0.001
0.06, 0.2, 1, 2, 10 or 100 (PLCs).
*
• To allow for normal mode (line frequency noise) rejection, you must
34410A/11A User's Guide
Measurement Function
dc voltage, dc current,
resistance, temperature
ac voltage, ac current
frequency / period
capacitance, continuity
check, diode check
These two additional settings are provided on Model 34411A only.
select an integral number of power line cycles (e.g. 1, 10 or 100 NPLC).
Selectable Measurement Time Parameter(s)
*
NPLC Values: 0.001
, 0.002
1, 2, 10, 100 (PLCs)
Aperture Values: settable from 300 3s to 1s
AC Filter Values: 3 Hz, 20 Hz, 200 Hz
AC Filter Values: 3 Hz, 20 Hz, 200 Hz
Gate Time Values: 0.001, 0.01, 0.1, 1 second
fixed measurement time
*
These two settings are provided on model 34411A only
**
Model 34411A is settable down to 100 3s.
Features and Functions
*
, 0.006 , 0.02, 0.06 , 0.2,
**
no selectable parameters
*
*
, 0.002
, 0.006, 0.02,
2
.
53