[SENSe:]VOLTage[:DC]:RESolution

Syntax
[SENSe:]VOLTage[:DC]:RESolution {<resolution>|MIN|MAX|DEF}
This command selects the measurement resolution for DC voltage
measurements. Specify the resolution in the same units as the selected
measurement function, not in number of digits.
[SENSe:]VOLTage[:DC]:RESolution? [{MIN|MAX}]
The query returns a numeric value that represents the DC voltage resolution.
Parameter
Item
resolution
Remarks
• For the <resolution> parameter, you can substitute MIN or MAX for a
• The instrument sets the resolution to 5½ digits after a Factory Reset (*RST
U3606B Programmer's Reference
Type
Numeric
numeric value. MIN selects the smallest value accepted, which gives the
highest resolution; MAX selects the largest value accepted, which gives the
least resolution.
command).

[SENSe:]VOLTage[:DC]:RESolution

Range of values
• MAX (4½ digit)
• MIN (5½ digit)|DEF
1 uV
10 uV
100 uV
1 mV
10 mV
100 mV
SENSe Subsystem
Default value
MIN
0.1 uV
1 uV
10 uV
100 uV
1 mV
10 mV
10